ADVANCED ENERGY 
ENVIRONMENTAL SPECIFICATIONS
Ambient Temperature
Operating
Storage
Transportation
Humidity
Atmospheric Pressure
Operating
Storage
Transportation
Coolant Requirements
Heat Removal
Temperature
Flow Rate
Pressure
Contaminates
1-4
5°C to +40°C (+41°F to +104°F)
-25°C to +55°C (-13°F to +131°F)
-25°C to +55°C (-13°F to +131°F)
15 to 85% relative humidity; non-condensing
≥ 745 mbar (≤ 2500 m (8,203') equivalent altitude)
≥ 585 mbar (≤ 4000 m (13,124') equivalent altitude)
≥ 480 mbar (≤ 5000 m (16,405') equivalent altitude)
11,270 BTU/hr (3300 W) at full rated RF output power
+5°C to +25°C (+41°F to +77°F) inlet temperature
15.1 liters (4 gallons) per minute minimum
100 psi (6.9 bars) maximum inlet pressure
The following specifications are recommended for the
water used to cool the RFG 5500:
• pH between 7 and 9
• total chlorine < 20 ppm
• total nitrate < 10 ppm
• total sulfate < 100 ppm
• total dissolved solids < 250 ppm
• total hardness expressed as calcium carbonate
equivalent less than 250 ppm
• specific resistivity of 2500 Ω per centimeter or
higher at 25°C
• total dissolved solids (TDS) as estimated by the
following:
TDS ≤
specific resistivity (in Ω/cm)
December 2000
640,000
P/N 5706045C