4.3.8 Humidity adjustment
For humidity adjustment, the measurement parameter of the connected probe
is adjusted to the reference value at the two standard adjustment points 11.3%
RH and 75.3% RH; any deviations between the measuring value and the
nominal value are minimised across the entire measuring range.
The Testo calibration set provides a reference value to calculate the offset for a
humidity adjustment.
Humidity adjustment is possible with the following probes:
Order number
Name
0636 9771
High-precision temperature-humidity probe with
Bluetooth
0636 9772
High-precision temperature-humidity probe, fixed cable
0636 9731
Temperature-humidity probe with Bluetooth
0636 9732
Temperature-humidity probe, fixed cable
0636 9775
Robust temperature-humidity probe for temperatures up
to +180 °C, fixed cable
testo 440 is switched on and the corresponding probe is connected.
Probe is already exposed to reference conditions (e.g. salt pot) for an
adequate adjustment time.
Adjustment time for humidity probe: at least 30 minutes.
You are in the Humidity Adjustment menu.
1
Select the relevant reference point of 11.3 or 75.3% RH with
2
Select the probe to be adjusted.
3
Use
to select Adjust and confirm with
An information window displays the remaining adjustment time and
the adjustment is carried out.
An information window displays the text Adjustment successful!.