Siemens SIREMOBIL Compact Service Instructions Manual page 26

Table of Contents

Advertisement

5 - 4
Fig. 1
Error examples:
Tube current exceeds the upper limit
- preset filament current too high
Fig. 1a
Tube current exceeds the lower limit
- preset filament current too low
Fig. 1b
SIREMOBIL Compact
System Manual
relevant time period
Register 5
RXR2-130.061.02
Rev. 03
Adjustments / Programs
kV-act (20 kV = 1.0 V)
mA-act (2.5 mA = 1.0 V)
6 ms
3 ms
6 ms
3 ms
Page 4 of 22
10.03
CS PS 24
14.5 mA
13 mA
11.5 mA
14.5 mA
13 mA
11.5 mA
Siemens AG
Medical Engineering