  • Page 1 ThermalTronix TT-1060MD-FTM Series Thermal Imaging Camera User Manual...
  • Page 2 General Information CONVENTIONAL Warning: The improper operation can cause the thermal imaging system the major damage。 Careful, attention, warning, danger: In the reminder operation should pay attention item Please read Note, hint, tip and consider: to operates the content and explanation. "WARNING"...
  • Page 3: Table Of Contents

    Table of contents 1. INTRODUCTION ..........................1 2. TECHNICAL SPECIFICATIONS ......................1 3. CONNECTIONS & BUTTONS DESCRIPTION ..................2 3.1 C ..........................2 ONNECTIONS 3.2 B ........................4 UTTONS DESCRIPTION 4. OPERATIONS ........................... 5 4.1 I ............................ 5 NTERFACE 4.2. M .......................
  • Page 4: Introduction

    1. Introduction Thank you for choosing the TT-1060MD-FTM thermal imaging camera (referred to hereafter as “camera” or “TT-1060MD-FTM”). The TT-1060MD-FTM uses the high sensitive UFPA detector. Simply connect the camera to a PC or a monitor and produce high-quality real-time images. Its low weight, ruggedness and compacted.
  • Page 5: Connections & Buttons Description

    3. Connections & Buttons Description 3.1 Connections Figure 3.1 Ethernet Port: For data and information transfers. Analog Video Output Connector: Connect large monitors for viewing. Power Supply Connector: Connect to power supply Serial Communication and I/O Port: Digital (TTL) output/input, YUN terrace communication and user-defined pins.
  • Page 6 Mode 1 RS485+ RS485- Spare Spare Mode 2 RS422 Output: RS485 and RS422 changeable GPIO 1,2,3 output: Inside framework GPIO 1,2,3 recommended connection: Item: VCC recommended:3.3V,5V,8V,12V。 RL recommended:10K VCE Output control pin ,We suggest adding amplifying circuit if the circuit drives high current component.
  • Page 7: Buttons Description

    3.2 Buttons description  S Button Press briefly to freeze an image  Press to leave freeze mode and go to live mode  Composite Composite Buttons consist of UP, DOWN, LEFT, RIGHT and CONFIRM Buttons buttons.  Press confirm button to display the menu system ...
  • Page 8: Operations

    4. Operations 4.1 Interface Figure 4.1 Callout Item Description Work mode To display current work mode Crosshair To display the position of a spot in the image Minimum temperature The minimum temperature of the color palette Color palette To show the scale of the thermal image Maximum temperature The maximum temperature of the color palette Result table...
  • Page 9: Menu Function Description

    4.2. Menu function description The menu system includes a number of different menu objects. This section describes the menu objects. Figure 4.2 4.2.1 Temperature measurement mode Figure 4.3  Press confirm button to display the pull up menu.  Point to Measurement item by pressing up/down buttons. ...
  • Page 10: Spot

    Figure 4.4  Press left/right button to seriatim change the content with the sequence of None, Add Spot, Horizontal, Vertical and Add Area. 4.2.1.1 Spot  Point to Measurement and then press the confirm button to display the Measurement dialog box.
  • Page 11: Horizontal

     Press left/right or up/down button to move the spot.  Press the confirm button to display the attribute dialog. Press up/down to change the select item and press the left/right to change the value of emissivity or set the reference temperature.
  • Page 12: Vertical

    Figure 4.6 4.2.1.3 Vertical  Point to Measurement and then press the confirm button to display the Measurement dialog box.  Press left/right button to select Vertical and then press the confirm button. A line and a curve will now appear on the screen. The measured temperature will be displayed in the result table in the top right corner of the screen.
  • Page 13: Area

    Figure 4.7 4.2.1.4 Area  Point to Measurement and then press the confirm button to display the Measurement dialog box.  Press left/right button to select Add Area and then press the confirm button. A temperature area will now appear on the screen. The measured temperature will be displayed in the result table in the top right corner of the screen.
  • Page 14: Level/Span

    Figure 4.8  Emissivity: Press the left/right button to change the emissivity.  Set Ref: Set the current temperature as reference temperature.  Area Type: To set the display temperature in result table, you can select maximum, average or minimum. ...
  • Page 15: Setup

    Symbols in the temperature scale, indicating (1) increasing span; (2) decreasing span;(3) decreasing level, and (4) increasing level 4.2.3 Setup This menu contains all the functions for changing the camera setup parameters. This section describes the parameter setup. Figure 4.10 4.2.3.1 Measure Set Figure 4.11...
  • Page 16: Measure Correct

    Point to Measure Set and then press the confirm button to display the Measuring Setup dialog box. Label Value comments Temp Range -20~180℃ To set the temperature range it could not be adjust by user. Optional Lens Standard To set the optional lens, set by manufacture Temp Unit ℃...
  • Page 17: Analysis

    Emissivi User-defi To set the emissivity Typical emissivity for a variety of common materials are listed in appendix 1.The value are meant to be (0.1~1) used only as a guide and can vary depending on many different factors. To set the system’s parameter correction(Usually set by Correct manufacture) To set the system’s temperature correction(Usually set by...
  • Page 18: Local Adapt

    the screen when the alarm is triggered. Isotherm Color 9 colors to be Sets an isothermal color. This setting command colors chosen all pixels with the setting temperature. The isotherm levels will be displayed in the temperature palette. Sets the temperature of the isothermal central point. Isotherm Temp User-defined Average distribute around the central isothermal...
  • Page 19: System Info

    4.2.3.5 System Info Figure 4.15 This prompt box shows information about Device ID, Software revision, and Camera Create Time etc. No changes can be made. 4.2.3.6 Fact Default Figure 4.16  Point to FactDefault and press the confirm button to display the Reset prompt box. ...
  • Page 20: Network Set

    4.2.3.7 Network Set Figure 4.17 Point to NetworkSet and then press the confirm button to display the network setting dialog box. the setting parameter will effective after restore the apparatus. Label Value Comments  Auto Yes/No Yes: The IP address will be distribute by DHCP services ...
  • Page 21: Client Software Operations

    5. Client software operations TT-1060MD-FTM_Client is accessional software you can use pc to control the camera by this software. Getting started Double click on to start up the software, with the interface shown as following: Figure 5.1...
  • Page 22: Graphical User Interface

    5.2 Graphical user interface 5.2.1 Interface and Connection description Interface work region Figure 5.2 There are four small work regions on the software interface, it’s said that four apparatus can be connect to the PC and operate by the software, and then operate one of the regions to control the corresponding apparatus.
  • Page 23: Interface Operation Description

    Figure 5.4  If the apparatus in the sleep mode will automatically awaken, the video will display on the screen if the connection is successful . Figure 5.5  Click the work region 1 to zoom in the region 1 to whole work region. Network transfer rate Apparatus...
  • Page 24 section 5.4. To analyze thermal image. it’s spare function in currently. Click to quit the software and then the TT-1060MD-FTM_Client dialog box will pop-out. (Right figure) Press Y to disconnect the apparatus. To clear all spots on the screen. For more information about clear spot see section 21.
  • Page 25 the frames according with the setting of “Frames”, and the interval of the adjacent frame decided by the setting of “Interval”. Click to start recording video. The record will be saved in compressed file. But the monitoring record could not be analyzed by the report software Stop the monitoring record.
  • Page 26: System Setting

    For spot temperature with its temperature value being measured in real-time. While the camera is in the monitoring mode, maximum four spots can be displayed with its temperature being shown respectively. Detailed description is as following: Example: Press button S2, and move the mouse onto the monitoring frame. Click on its let key on the position you appointed, a spot with No.
  • Page 27 IP address:Set IP address of the apparatus. Note: the IP address setup must same as the current apparatus Figure 5.8 To set the emissivity of spots, line and areas .To select the display temperature type of areas. Figure 5.9 Click button to change the path.
  • Page 28 Item Optional Comments Video Type PAL/NTSC Analog video output Video Mode Auto/Manua Brightness, contrast adjustment mode Temp on Image Yes/No To show or hide the temperature result and palette. ISO Temp User-defined Isotherm temperature. Reference the 4.2.3.3 ISO Hight User-defined Isotherm height Reference the 4.2.3.3 ISO Color User-defined...
  • Page 29 Figure 5.12 Item Value Comments Remote Alarm Yes/No Select to active or inactive remote alarm Local Alarm Yes/No Select to active or inactive local alarm Alarm Temp User-defined To set the alarm temperature Alarm Color Optional To select alarm color Figure 5.13 Item Parameter...
  • Page 30: Video Play Software

    Figure 5.14 Click button to change the MEPG4 video saved path. Figure 5.15  Protocol:To select the communication protocol with Yun terrace. Optional: PELCO-D and YAAN  Addr:Spare address 5.4. Video play software Figure 5.16  Click to display a windows dialog box, select the file that you want and then click open to play the video.
  • Page 31 Figure 5.17  Click to play the video in full screen.  To control the play speed  These buttons are used for controlling video play.  To record a section of the video. You can set the save path via operate the setting dialog box.
  • Page 32: Trouble Shooting

    6. Trouble shooting If you have some trouble like following, please compare with the following contents to ravel out the problem, otherwise please switch off the power supply and contact with the manufactory. Problem Problem reason and solution One TT-1060MD-FTM camera only allow The PC could not connect with the to connect one visitor at a time.