Summary of Contents for Intellisystem ThermalTronix TT-607FG-HTI
Page 1
INFRARED GAS DETECTION THERMAL IMAGING CAMERA ThermalTronix TT-607FG-HTI User Manual...
Page 2
Warnings and Precautions · Infrared gas leak temperature imager uses a very sensitive thermal sensor, so in any case (power on or off) shall not be directed at the device lens strong radiation sources (such as the sun, direct laser beam or reflection, etc.) , Otherwise it will cause permanent damage to the Infrared Gas Leak Detector Imager.
FACTORY SETTING PARAMETER TABLE ............... 20 1 Introduction Thank you for choosing TT-607FG-HTI infrared gas leak detection temperature imager. TT-607FG-HTI infrared gas leak detection temperature imager is infrared system, long-wave infrared quantum well focal plane array detector, electronic processing systems and image processing software, the infrared radiation measurement object is converted to visual thermal images, to long-range Detection of sulfur hexafluoride gas and other trace leakage point, and quantitatively shows the target surface temperature, is a fast, accurate, non-contact detection of live equipment and other micro-leakage and...
Page 7
Name Function To freeze or activate the thermal image. Twice press to save photo Freeze/Save in imaging. There will be a “*” mark besides the Intellisystem (Mark S) Logo when the imaging is feezed. Calibration (Mark Long press to make calibration. One press to choose Color,...
focus (Mark C) SD Card For SD card to save file. Microphone For voice recording LCD Screen To see imaging 2.2 Interface Pic 2-2-1 TT-607FG-HTI Interface at back side. Mini USB Data transfer. Power supply DC10~15V Audio output HDMI Video output...
3 Basic Operation. 1.3 Mounting Accessories. 3.1.1 Battery 3.1.1.1 Install /Uninstall Install:push the battery into the camera as this photo. Uninstall:Press the battery latch to the right, then pull out the battery. Note: The device can only use the standard battery, or the battery may be stuck, the voltage does not damage the device and so on.
3.1.1.2 Battery safety use common sense 20 ℃ ∽ 20 ℃ 1、 The battery should be kept as far as possible - ambient temperature, the battery during storage due to the existence of a small amount of self-discharge phenomenon, in order to avoid the battery during storage may have the impact of over-discharge battery capacity, the battery should be fully charged Save, and at regular intervals for a charge.
2.3 Quick Guide. 3.2.1 Get infrared imaging Load the battery or power supply. Connect with the monitor if needed. Long press the power button and wait for the camera imaging. Open the lens cap and towards the imager to the target object. Manual/Auto focus the lens to get the best imaging quality.
3.2.3 Save photo/Voice note. Manu save:Menu->Document->Save, then press OK button. We suggest to freeze the imaging before save photo. Fast save:Press “S” button to freeze the imaging, then press “S” button to save photo. Voice not:choose the microphone icon when imaging play back to record the voice note.
Pic 4-1-1 Main interface Name Description Back Back to real time imaging Time Time Save/Electrical zoon/Battery Show Save/Electrical zoon/Battery states Specific radiation coefficient Specific radiation coefficient Record video Hot key to record video Color Palette Choose different color palette Take photo Take photo Auto/Manual Switch auto/manual calibration...
4.2.1.1 Measurement settings Temperature range: - for temperature measurement, Observed、 40/180℃、100/500℃ Shake mode for gas detection. Extend lens:need to change if the camera fixed with any extend lens. Temperature Units: The currently available temperature units are ℃, K. Ambient Temperature: Infrared Gas Leak Detector When the imager is turned on, it automatically detects the ambient temperature.
the temperature of the measured object. 4.2.1.3 Analysis settings Temperature alarm: This setting is valid only in the area temperature measurement. When the temperature alarm is set to "On", if the target temperature in the temperature measurement area exceeds or falls below the alarm temperature, the infrared gas leak detector Automatic sound and color alarm.
4.2.1.5 System settings Language: Set the infrared gas leak detection temperature imager using the language. Auto Calibration: Set the auto calibration interval to get better thermal image and improve the accuracy of temperature measurement. Electronic magnification: electronic magnification, there are three options: X1, X2 and X4. Laser: Set the laser on and off.
4.2.1.8 Factory settings Reset to factory initial state. 4.2.2 Document menu 4.2.2.1 Open a) Playback jpg Record voice, 40secs at mots Save Play b) Playback mpeg4 (m4v) c) Playback raw (DLV original data)
4.2.2.2 Save 0 second means no auto save. 4.2.2.3 Record Note: mpeg4 meet standard ISO14496-2, for play in PC. DLV for analysis software. 4.2.2.4 SD Choice. Choose from SD1 and SD2. 4.2.2.5 SD unload Please remember to unload the SD card here before remove the SD card.
4.2.2.6 Delete file. 4.2.2.7 format 4.2.3 Measurement mode Add point: Add the temperature measurement point Add line: Add the line temperature line Add area: Add the regional temperature measurement area Clear: Cancel all temperature Auxiliary item Note: Only Infrared mode is active.
Page 21
5 Factory setting parameter table Temperature setting table Name Number Specific radiation coefficient 0.96 Correct the temperature Distance Humidity Analysis Settings Table Name Number Temperature range -40/180°C Area temperature measurement Highest temperature Alarm Alarm temperature 50.0°C Alarm color Auto Reference temperature Reference temperature Value 0.1°C Color Palette...
Page 22
Appendix Specific Emissivity of Materials Material Surface Temperature °C Emissivity ( ε ) Non- oxidized 0.20 Aluminum Oxidized 0.55 Burnished to Brown 0.40 Brass Dull luster 0.22 Oxidized 0.61 Copper Seriously Oxidized 0.78 Oxidized 0.74 Iron Rusty 0.65 Oxidized 0.64 Cast iron Non- Oxidized 0.21...
Page 23
Technical Specifications ThermalTronix Item TT-607FG-HTI Detector Type Cooled QWIP detector Detector Array size 320*256 Field of View/min 14.5°×10.8°/0.5m or 24°×18°/0.3m focus distance Spatial resolution 0.79mrad or 1.13mrad NETD <0.025℃ Image Characteristics Frame rate 60Hz Focus Automatic/Manual/electric focusing Zoom 9.8~11.2μm,Summit 10.55μm 320million pixel CMOS,AF View finder High resolution 0.6’...
Page 24
Storage Card 8G SD card,could up to 32G Manual/automatic single-frame image storage, continuous Storage mode visible, infrared video recording Thermal/CCD, CCD thermal image automatic association of Storage mode the corresponding Image Save Single frame JPEG, 14-bit thermal image with measurement data Video MPEG-4 or 14-bit thermal image with measurement data Single frame...