In the 【Menu】 interface, select "Exit" to return to 【Main】interface. If the measurement is not completed
before entering 【Main】interface, it will return to the 【Examination】 interface.
6.1.5 Repeat the measurement
The measurement of the device can be repeated. Long press the "Repeat" button to enter the 【Examination】
interface. When the memory is full, it will display【The memory is full】, as shown in the Picture 27. If you select
"Yes", it will enter the【Delete data】interface; if you select "No", it will enter the【Menu】interface.
6.1.6 Charging
There are two ways of charging:
1) Connect the device to the computer with a USB cable.
2) Connect the device to the power source with the power adapter.
To charge the device, connect it to a network from which you can easily disconnect. When charging is
complete, unplug the power adapter.
6.1.7 Data downloading
The device automatically transfers the measurement data on the monitor or the tablet with installed IDIS2GO
software.
Attention!
Before use, check the device and make sure that it can work properly.
A lithium rechargeable battery is used.
It is recommended that the device be used indoors.
Excessive ambient light may affect the measurement accuracy. This includes a fluorescent lamp, double
ruby light, infrared heater, direct sunlight and more.
Intense subject activity or extreme electrosurgical interference can also affect accuracy.
Please clean and disinfect the device after use according to the User manual (7.1).
Telehealth medical diagnostic system IDIS2GO
Picture 27
13