Picture 21: Time settings interface
d) Calibration
In the【Settings】interface select "Time" to enter the【Calibrate parameter】interface, as shown in the
Picture 23. Select 2L or 3L according to the volume of the syringe, then go to the "Calibration" interface, as shown
in the Picture 24.
Picture 23: Calibration interface
Under the【Calibration】, interface, press the syringe once, the device will show "REPEAT", then press the
syringe again. After two correct continuous operations, the calibration will be successful and the device will show
"OK". Finally, the interface will switch to the previous interface before calibration (previous interface: if the device
is calibrated after the measurement is completed, it will return to the 【Settings】 interface; if it is calibrated
before the measurement is completed, it will return to the 【Examination】 interface).
If "Error! Please repeat" is displayed on the device, it means something is wrong with the operation, please
repeat the calibration until success. If "Select the correct volume" appears on the device, please confirm the syringe
volume and calibration selection, and then repeat the calibration until successful. If you need to stop the calibration,
just press the "Confirm" button to go to the previous interface before calibration.
In the【Calibrate parameter】interface select "Adjust" to enter the【Adjust】interface, as shown in the
Picture 25. Press the Up or Down buttons to change the value (long press available), then press the Confirm button
to return to the【Confirm adjustment】interface, as shown in the Picture 26. Selecting "Yes" will save the
configured value, selecting "No" will cancel the setting, and the device will return to the【Calibration】interface.
Notation
The set values determine the accuracy of the measurement, please DO NOT change them randomly. After
replacing the turbine, calibration is used to enter the parameters of the new turbine, which guarantees the
accuracy of measurements after replacing the turbine.
Picture 25: Calibration interface
In the 【Calibration】interface select "Exit" to return to the【Settings】interface.
e) About
In the【Settings】interface select "About" to enter the【About】interface. The user can view the device
name and software version. Click the "Confirm" button to return to the【Settings】interface.
f) Exit
In the【Settings】interface select "Exit" to return to the【Menu】interface.
Power off
In the 【Menu】interface select "Power off", the device will turn off.
Notation. If the operation is not taken within 1 minute, the device will turn off automatically.
Exit
Telehealth medical diagnostic system IDIS2GO
Picture 22: Minute setting interface
Picture 24: Calibration interface
Picture 26: Adjustment confirmation interface