Instructions for use PIEZOlux (0.571.5211 - 1.001.8522) Contents Contents Contents User instructions User guide 1.1.1 Abbreviations 1.1.2 Symbols 1.1.3 Target group Service Transportation and storage Safety Description of safety instructions 2.1.1 Warning symbol 2.1.2 Structure 2.1.3 Description of danger levels Purpose –...
Electro magnetic compatibility 1.1.2 Symbols Refer to chapter Safety/Warning Important information for users and technicians KaVo products with this symbol can be sterilised in autoclaves up to 138°C. Action request Material number CE mark according to EC Directive 93/42 for medical devices Disposal instructions, proper use Read and note the content of accompanying documents.
Service hotline: +49 (0) 7351 56-2500 Service.Einrichtungen@kavo.com Please indicate the product serial number in all requests. Additional information can be obtained at: www.kavo.comAdditional information can be obtained at: www.kavo.com 1.3 Transportation and storage CAUTION It is hazardous to start up the medical device after it has been stored refrigerated.
KaVo product for the intended purpose are to be applied and complied with. This KaVo product is intended only for use in the field of dentistry. The product may not be used for a purpose for which it was not intended.
The PIEZOlux Paro tips are used in periodontology (gentle cleaning). PIEZOlux cem tips are used for the insertion of inlays (in the semi-liquid composite and quickly liquefies). Note KaVo PIEZOlux tips can also be used with the KaVo SONOsoft. 2.3 Safety 2.3.1 Product-specific safety instructions CAUTION Sharp-edged tip.
Instructions for use PIEZOlux (0.571.5211 - 1.001.8522) 3 First use | 3.1 Exchanging tips 3 First use Note Before the first use of the PIEZOlux handpiece, sterilise the removable sleeve. 3.1 Exchanging tips 3.1.1 Screw out the tip ▶ Place the torque wrench on the tip, and unscrew it counterclockwise. 3.1.2 Screw in the tip Note Before initial use, sterilise the tips.
4.1 Using the PIEZOlux See also: Instructions for use for the KaVo treatment unit The KaVo ZEG PIEZOlux can gently remove calculus and plaque as well as excess cement from natural teeth without pain at a vibrational frequency of 30 kHz.
Improper service and care can cause premature wear and malfunctioning. KaVo only guarantees that its products will function properly when disinfectants are used that are listed by KaVo since they were tested for proper use on KaVo prod‐ ucts. Infection protection For treatments that are a "critical treatment according to RKI,"...
Instructions for use PIEZOlux (0.571.5211 - 1.001.8522) 5 Preparation methods DIN EN ISO 17664 | 5.1 Cleaning ▶ Switch main device switch off. ▶ Clean the instrument and instrument with a moist, clean cotton cloth until no residual soiling can be seen. Automatic cleaning Not applicable.
▶ Do not place the medical device in disinfectant baths. 5.2.1 Disinfecting the instrument Manual disinfection KaVo recommends the following products based on material compatibility. The mi‐ crobiological efficacy must be ensured by the disinfectant manufacturer. Permissible disinfectants: ▪ AF liquid ▪...
5.2.3 Disinfecting the ZEG tips Manual disinfection KaVo recommends the following products based on material compatibility. The mi‐ crobiological efficacy must be ensured by the disinfectant manufacturer. Permissible disinfectants: ▪ AF liquid ▪...
Instructions for use PIEZOlux (0.571.5211 - 1.001.8522) 5 Preparation methods DIN EN ISO 17664 | 5.3 Sterilisation 5.3.1 Sterilising the grip sleeve Requirement Remove the grip sleeve from the handpiece. ▶ Seal the grip sleeve in sterile goods packaging. ▶ Sterilise the grip sleeve an autoclave. Depending on the device, select a suitable method from the following sterilisation methods.
Instructions for use PIEZOlux (0.571.5211 - 1.001.8522) 5 Preparation methods DIN EN ISO 17664 | 5.4 Service and Care Autoclave using the gravitational method: sterilise for at least 10 min. at 134°C ± 1°C (273°F ± 1.8°F) Autoclave using the gravitational for at least 60 min. at 121°C ± 1°C (250°F ± 1.8°F) Follow the manufacturer's instructions for use.
Instructions for use PIEZOlux (0.571.5211 - 1.001.8522) 5 Preparation methods DIN EN ISO 17664 | 5.5 Control and function test 5.5.1 Check the calculus remover tips with a measuring card Wear to the calculus remover tips from frequent use. Check the permissible length using the measuring card Mat.
Instructions for use PIEZOlux (0.571.5211 - 1.001.8522) 6 Tools 6 Tools Note Available from dental suppliers: Designation Material number PIEZOlux Scale tip No. 9 Mat. no. 0.571.5822 PIEZOlux Scaler tip No. 10 Mat. no. 0.571.5832 PIEZOlux Scaler tip No. 11 Mat.
Need help?
Do you have a question about the PIEZOlux 0.571.5211 and is the answer not in the manual?
Questions and answers