Appendix

To Use Auxiliary Suction Ports (optional)

The auxiliary suction ports allow connection to a hospital wall suction
regulator if an alternate suction source is required.
WARNINGS:
The rover CANNOT regulate hospital wall suction and CANNOT
display suction levels above 570 mm-Hg. ALWAYS use a hospital
wall suction regulator to control and monitor suction levels when
using the auxiliary suction ports. Failure to comply could result in
serious injury or death.
ALWAYS make sure rover power is ON when collecting fluid
waste. The rover can only detect full canisters if the rover is ON.
If the rover is OFF, biohazard waste leakage or loss of suction
can occur.
NOTES:
The rover is equipped with two auxiliary suction ports for each
canister. Connecting both auxiliary ports of a canister to the
hospital wall suction regulator will increase airflow and improve
suction performance compared to connecting only one auxiliary
port of a canister.
If the ports are used, electrical power must be applied to the
rover. The rover will not regulate the vacuum levels but will display
vacuum and volume values and provide an audible indication if
the canister fluid levels exceed their pre-defined limits.
ALWAYS make sure the SUCTION indicator is OFF (not
illuminated). If illuminated, push the SUCTION button to turn the
indicator OFF.
Applying maximum suction levels up to 760 mm-Hg will not
damage the rover.
1. Verify the rover is connected to facility power and push the power
switch to the ON position (Figure 45).
Figure 45 To Connect Power
www.stryker.com
Printed Date: Feb 24, 2020 18:26:59 GMT
0000016914, Rev. S Effective Date: Mar 11, 2016 9:46:10 AM
2. Read the WARNING message on the control panel display, then
push the control panel display button next to OK to access the
CONTROL screen (Figure 46).
3. Attach suction tubing between the hospital wall suction regulator
and the auxiliary suction ports on the rover (Figure 47).
4. Apply and regulate the suction level provided to the rover from the
hospital wall suction regulator.
1
WARNING
HIGH SUCTION DEVICE
DO NOT CONNECT DIRECTLY
TO CHEST TUBES OR OTHER
CLOSED WOUND DRAINS
Figure 46 High Suction Device Warning
4
3
Figure 47 To Connect Auxiliary Suction Ports
EN
OK
2
39