Stryker System 6 6203-000-000 Instructions For Use Manual
Stryker System 6 6203-000-000 Instructions For Use Manual

Stryker System 6 6203-000-000 Instructions For Use Manual

Single trigger rotary handpiece

Advertisement

Quick Links

Dimensions:
NOTE: This portion of the
6 inch (width) x 6 inch
document should not appear
Booklet
on produced Labels or IFUs.
Stryker Instruments
Print Location:
(269) 323-7700
(800) 253-3210
Print Center
DSGN-fm-50366 rev. None
Effective: July 1, 2008
Color/Material/Finish:
Color Graphics on White Background
20# Bond or Equivalent
Suppliers/Services:
N/A
Label Stock:
N/A
Description/Type:
Instructions For Use
Part Number:
6203-001-700
Page 1 of 1
Rev.
E

Advertisement

Table of Contents
loading

Summary of Contents for Stryker System 6 6203-000-000

  • Page 1 6 inch (width) x 6 inch Color Graphics on White Background document should not appear Booklet 20# Bond or Equivalent Description/Type: on produced Labels or IFUs. Instructions For Use Stryker Instruments Print Location: Suppliers/Services: Part Number: Rev. (269) 323-7700 (800) 253-3210 Print Center 6203-001-700 DSGN-fm-50366 rev.
  • Page 2 System 6 Single Trigger Rotary Handpiece 6203-000-000 Instructions For Use ENGLISH (EN) 2015-03 6203-001-700 Rev-E www.stryker.com...
  • Page 3: Indications For Use

    Stryker sales representative or call Stryker customer ▪ Upon initial receipt and before each use, clean service. Outside the US, contact your nearest Stryker and sterilize the equipment as indicated. See subsidiary. the care instructions manual supplied with the Indications For Use equipment.
  • Page 4 NOTE: For a complete list of accessories, contact mixed with air, oxygen, or nitrous oxide. your Stryker sales representative or call Stryker ▪ Take special precautions regarding customer service. Outside the US, contact your electromagnetic compatibility (EMC) when nearest Stryker subsidiary.
  • Page 5 Applied Parts - The distal end of the handpiece and the attachment (as defined by the standards listed in the Specifications section under Product Safety Certification). Shifter Knob Applied Parts F/R/Safe Control Attachment Accessory Interface Actuating Collar Battery Latch Trigger Switch Battery Pack www.stryker.com...
  • Page 6 See the Symbol Definition Chart supplied with the equipment. SYMBOL DEFINITION SYMBOL DEFINITION Forward Mode DRILL Reverse Mode REAM General warning sign Per European Union Directive 2012/19/ EU, product must be collected separately. Do not dispose of as unsorted municipal waste. Contact local distributor for disposal information. www.stryker.com...
  • Page 7: To Install Battery Pack

    Large or Small Battery Pack only (REF 6215-000- position. 000 or REF 6212-000-000). These battery packs can be charged in the Stryker System 6 Battery 2. To install an attachment, align the driveshaft of Charger (REF 6110-120-000) or the Universal the attachment with the handpiece spindle.
  • Page 8 2. Squeeze the pressure sensitive trigger for variable ▪ Ensure the F/R/Safe control DOES NOT change speed operation. position, for example, forward to reverse, while the 3. Slide the F/R/Safe control to the safe mode handpiece is operating. position when you are finished operating the handpiece. www.stryker.com...
  • Page 9: Troubleshooting

    Troubleshooting WARNING: DO NOT disassemble or service this equipment. NOTE: For service, contact your Stryker sales representative or call Stryker customer service. Outside the US, contact your nearest Stryker subsidiary. PROBLEM CAUSE...
  • Page 10 Sporadic electrical Electrical noise is present. Turn off all electrical equipment not in use in interference is the operating room. experienced. Relocate electrical equipment; increase spatial distance. Plug operating room equipment into different operating room outlets. www.stryker.com...
  • Page 11: Care Instructions

    Type BF Applied Part Maximum Temperature Less than 124 °F [51 °C] (Maximum surface temperature as tested to the standards of Applied Parts: listed under Product Safety Certification.) Power Supply: Internally Powered 9.6 V (Direct current) Ingress Protection: IPX0 Ordinary Equipment www.stryker.com...
  • Page 12: Product Safety Certification

    Edition; Revisions through and including April 26, 2006 European Committee for Electrotechnical Standardization (CENELEC) EN 60601-1:2006, Medical Electrical Equipment — Part 1: General Requirements for Basic Safety and Essential Performance; IEC Corrigendum 1 (2006); IEC Corrigendum 2 (2007); CENELEC Corrigendum (2010) www.stryker.com...
  • Page 13 000-000) is suitable for use in all establishments other than Harmonic emissions domestic and those directly connected to the public low- IEC 61000-3-2 voltage power supply network that supplies buildings used Voltage fluctuations/flicker emissions for domestic purposes. IEC 61000-3-3 www.stryker.com...
  • Page 14 Power frequency magnetic fields should be at levels characteristic of a typical 3 A/m 3 A/m magnetic field location in a typical commercial or IEC 61000-4-8 hospital environment. NOTE: U is the alternating current mains voltage prior to application of the test level. www.stryker.com...
  • Page 15 If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the System 6 Single Trigger Rotary Handpiece (REF 6203-000-000). Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. www.stryker.com...
  • Page 16 NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. www.stryker.com...
  • Page 17 ES/DE/FR/IT/NL 6203-001-710 JA/ZH/KO 6203-001-720 SV/DA/FI/PT/NO 6203-001-730 PL/EL/TR 6203-001-750 Stryker Instruments 4100 E. Milham Kalamazoo, Michigan (USA) 49001 1-269-323-7700 1-800-253-3210 2015-03 6203-001-700 Rev-E www.stryker.com...

Table of Contents