1280 Enterprise™ Series

Enabling Bootloader

Product: 1280 Enterprise Series indicator with firmware v1.10
Issue: Display locks in "Booting" process when attempting to update firmware
Solution: Manually enable bootloader with configuration and reset switches
Manuals and additional resources are available from the Rice Lake Weighing Systems website at
Warranty information can be found on the website at

Procedure

When upgrading the 1280 Enterprise Series with firmware v1.10, the bootloader may need to be manually enabled if the display
is locked at the "Booting" step.
The bootloader can be manually enabled by observing the following sequence during power up:
1. Press and hold the configuration switch
2. Power on the indicator and continue to hold down the configuration switch while the system is loading.
3. Press the reset switch
The fourth dot only appears momentarily. If the reset switch is not enabled precisely on the 4th dot, the display may
become locked at the "Booting" step again. It may be necessary to repeat this procedure until successful.

Access to Switches

Access to Configuration Switch

The configuration switch can be accessed by removing the fillister screw in the bottom of the 1280 enclosure and inserting a
non-conductive object into the hole to press the button.
November 6, 2019
(Figure
(Figure 4 on page
2) while the splash screen displays the fourth dot in the opening sequence.
Figure 1. Location of Configuration Switch
www.ricelake.com/warranties
1).
Access to
Configuration Switch
Fillister Screw
www.ricelake.com
PN 198132

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Enterprise 1280 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Rice Lake Enterprise 1280 Series

  • Page 1 Issue: Display locks in “Booting” process when attempting to update firmware Solution: Manually enable bootloader with configuration and reset switches Manuals and additional resources are available from the Rice Lake Weighing Systems website at www.ricelake.com Warranty information can be found on the website at www.ricelake.com/warranties...
  • Page 2 Specifications subject to change without notice. Rice Lake Weighing Systems is an ISO 9001 registered company. 230 W. Coleman St. • Rice Lake, WI 54868 • USA U.S. 800-472-6703 • Canada/Mexico 800-321-6703 • International 715-234-9171 • Europe +31 (0)26 472 1319...

Table of Contents