FIRST RESPONDER - USER HANDBOOK GMI welcomes comments on all our publications. Your comments can be of great value in helping us to improve our customer publications. Please send any comments that you have to our Sales Department at GMI. Contact details are provided inside the back cover of this handbook.
Page 3
Ltd is prohibited. Reverse engineering is not permitted. LIABILITY Every care has been taken in the preparation of this document, but GMI Ltd do not accept any responsibility for errors or omissions and their consequences. Information in this document is subject to change without notice. This document does not constitute a specification or basis for a contract.
Page 4
The combustion chamber is a flameproof assembly and must not be opened in the presence of a flammable atmosphere. First Responder instruments are certified as EEx iad IIB T3 (-20 C < Tamb < 50 C). BAS03ATEX2448X II 2 G.
Page 17
INTRODUCTION The First Responder has a variety of user configurable options. This handbook details the default configuration, with possible options detailed in italic text.
Page 22
FIRST RESPONDER - USER HANDBOOK Fig. 2-3 PPM Range Fig. 2-4 LEL Range...
Page 23
GENERAL INFORMATION Fig. 2-5 Volume Gas Range...
Page 24
FIRST RESPONDER - USER HANDBOOK Fig. 2-6 Carbon Monoxide Range Alarms It is the responsibility of the user to ensure that the alarm levels, where set in the instrument, are appropriate for the safe operation and legal requirements for the country / industry in which the unit is being used.
Page 25
GENERAL INFORMATION Datalogging Fig. 2-7 Automatic Datalogging Enabled...
Page 26
FIRST RESPONDER - USER HANDBOOK Construction...
Page 31
OPERATING INSTRUCTIONS Instrument Features Fig. 3-1 Instrument Front Plate...
Page 32
FIRST RESPONDER - USER HANDBOOK Switch Panel PUMP AUTO RANGE ZERO ALARM ALARM ACK. Fig. 3-2 Switch Panel SINGLE PRESS (functions in red) DOUBLE PRESS (functions in blue) SINGLE PRESS and HOLD (to turn instrument ON)
Page 33
OPERATION Connection of Probe Fig. 3-3 Standard Probe and Handle Flexible Probe 8-Pin instrument connector plug Probe Sensor Fig. 3-4 Semiconductor Sensor Flexi Probe...
Page 34
FIRST RESPONDER - USER HANDBOOK 8-Pin Instrument Connector Plug Sample Line Probe Sensor Fig. 3-5 Semiconductor Survey Bellows Probe...
Page 35
OPERATION Connection for Connection for Probe Sample Line 8-Pin Probe Connector Plug Fig. 3-6 Probe Connection Measure Operation...
Page 36
FIRST RESPONDER - USER HANDBOOK Current detection reading Fig. 3-7 Warm Up At the end of warm-up and before the first Responder instrument is ready for measuring, the instrument will indicate on the display when the next calibration is due.
Page 37
OPERATION The re-calibration interval pre-set for all First Responder instruments is twelve (12) months. This period can be altered, however, you should ensure that the instrument is always within its calibration period prior to use. Note:...
Page 38
FIRST RESPONDER - USER HANDBOOK auto datalogging is disabled, the STORE flag being displayed for 1 second.
Page 54
FIRST RESPONDER - USER HANDBOOK Caution: Under circumstances should rechargeable batteries be fitted in place of alkaline. Filter Replacement Caution: The instrument should never be switched on without suitable filters installed.
Page 56
FIRST RESPONDER - USER HANDBOOK Fig. 5-8 Probe Handle and In-line Filter Assembly Caution: The instrument should never be switched on without suitable filters installed.
Page 58
FIRST RESPONDER - USER HANDBOOK In-line Water Filter Water Filter Housing Assembly Hydrophobic Filter Probe Sealing Washer yellow Probe Washer label Fig. 5-10 Water Filter Assembly 2) Remove Hydrophobic Filter, then discard.
Page 59
OPERATOR MAINTENANCE Probe Sealing Washer Fig. 5-11 Semiconductor Survey Bellows Probe Filter Assembly Caution: The instrument should never be switched on without suitable filters installed.
Page 61
OPERATOR MAINTENANCE Water Filter (Accessory) Water Filter Housing Assembly Hydrophobic Filter Assembly Probe Washer Fig. 5-13 Water Filter Assembly...
Page 62
FIRST RESPONDER - USER HANDBOOK Caution: The instrument should never be switched on without suitable filters installed. Coalescing Filter Assembly...
Page 63
OPERATOR MAINTENANCE Caution: The probe should never be used without suitable filters installed. Fig. 5-15 Semiconductor Sensor Flexi Probe...
Page 64
FIRST RESPONDER - USER HANDBOOK Fig. 5-16 Filter Replacement...
Page 77
FIELD CALIBRATION PUMP AUTO RANGE ZERO ALARM ACK. ALARM Fig. B-1 Switch Functions...
Page 78
FIRST RESPONDER - USER HANDBOOK Fig. B-2 Instrument Overlay Card Selectable Ranges in FCM...
Page 79
FIELD CALIBRATION Entering FCM Note: Fig. B-3 Field Calibration Display...
Page 80
FIRST RESPONDER - USER HANDBOOK Press and Single Press Double Press Hold Switch 1 Toggle Pump Exit CAL Mode Enter SPAN Mode On / Off Switch 2 Zero Current Save CAL Data Next Range Range Switch 3 Increase Exit SPAN –...
Page 81
FIELD CALIBRATION Field Calibration Procedure...
Page 82
FIRST RESPONDER - USER HANDBOOK Fig. B-5 Connecting Gas...
Page 83
FIELD CALIBRATION Fig. B-6 SPAN Mode Display...
Page 84
FIRST RESPONDER - USER HANDBOOK Fig. B-7 50% LEL Display Quitting FCM...
Page 104