CONFIGURATION & FIELD CALIBRATION Issue 1 17/09/05 Part Number: 67160 GMI welcomes comments on all our publications. Your comments can be of great value in helping us to improve our customer publications. Please send any comments that you have to our Sales Department at GMI.
Page 3
Ltd is prohibited. Reverse engineering is not permitted. LIABILITY Every care has been taken in the preparation of this document, but GMI Ltd do not accept any responsibility for errors or omissions and their consequences. Information in this document is subject to change without notice. This document does not constitute a specification or basis for a contract.
SWITCH INSTRUMENT ON / OFF in fresh air Fig. 1.1 Switch ON...
Page 12
CONFIGURATION & FIELD CALIBRATION Fig. 1.2 GT Model 1.1 INSTRUMENT IDENTIFICATION Fig. 1.3 Instrument Identification...
Page 13
SWITCH INSTRUMENT ON / OFF 1.2 BATTERY STATUS 100% Fig. 1.4 Battery Capacity 1.3 TIME AND DATE Fig. 1.5 Time and Date...
Page 14
CONFIGURATION & FIELD CALIBRATION 1.4 CALIBRATION DUE DATE 1. Cal Due Date message is not displayed. 2. Cal Due Date and overdue date messages are displayed Fig. 1.6 Calibration Due Date Fig. 1.7 Calibration Overdue...
Page 15
SWITCH INSTRUMENT ON / OFF 3. Cal Due Date message is displayed with user acknowledge if overdue (Default) Fig. 1.8 Calibration Due Date Fig. 1.9 Calibration Overdue...
Page 16
CONFIGURATION & FIELD CALIBRATION Fig. 1.10 Switch OFF...
Page 17
SWITCH INSTRUMENT ON / OFF 4. Cal Due Date message is displayed with user acknowledge for extended period option, if overdue Fig. 1.11 Calibration Due Date Fig. 1.12 Calibration Overdue...
Page 18
CONFIGURATION & FIELD CALIBRATION Fig. 1.13 Switch OFF...
Page 19
SWITCH INSTRUMENT ON / OFF 5. Cal Due Date message is displayed with user shut-down if overdue Fig. 1.14 Calibration Due Date Fig. 1.15 Switch OFF...
Page 20
CONFIGURATION & FIELD CALIBRATION 1.5 SERVICE DUE DATE 1. Service Due Date message is not displayed (Default). 2. Service Due Date and overdue date messages are displayed Fig. 1.16 Service Due Date...
Page 21
SWITCH INSTRUMENT ON / OFF Fig. 1.17 Service Overdue 3. Service Due Date message is displayed with user acknowledge if overdue Fig. 1.18 Service Due Date...
Page 22
CONFIGURATION & FIELD CALIBRATION Fig. 1.19 Service Overdue Fig. 1.20 Switch OFF...
Page 23
SWITCH INSTRUMENT ON / OFF 4. Service Due Date message is displayed with user acknowledge for extended period option, if overdue Fig. 1.21 Service Due Date Fig. 1.22 Service Overdue...
Page 24
CONFIGURATION & FIELD CALIBRATION Fig. 1.23 Switch OFF...
Page 25
SWITCH INSTRUMENT ON / OFF 5. Service Due Date message is displayed with user shut-down if overdue Fig. 1.24 Calibration Due Date Fig. 1.25 Switch OFF...
Page 26
CONFIGURATION & FIELD CALIBRATION Warm-up Complete The following configurable options are available: a) To start up in any selected other operational mode. b ) To start up in the mode last used. 1.6 SWITCH THE INSTRUMENT OFF SWITCH INSTRUMENT Fig.
Page 27
SWITCH INSTRUMENT ON / OFF Fig. 1.27 OFF sequence...
Page 32
CONFIGURATION & FIELD CALIBRATION Fig. 2.5 Configuration & Field Calibration Menu 2.2 ENTER CALIBRATION MENU Fig. 2.6 Calibrate Highlighted ...
Page 33
FIELD CALIBRATION Fig. 2.7 Enter Calibration Code Fig. 2.8 Calibration Disabled 2.2.1 Enter Calibration Code ...
Page 34
CONFIGURATION & FIELD CALIBRATION 2.3 ZERO ALL RANGES All sensors must be zeroed before the calibration can commence. Make sure that the instrument is in clean air. Fig. 2.9 Before Zeroing To zero all ranges: • Press ZERO Note: If an oxygen sensor is fitted, the sensor is zeroed to 20.9% Oxygen.
Page 35
FIELD CALIBRATION 2.4 CALIBRATE Fig. 2.11 Calibration Screen 2.4.1 Ranges 2.4.2 Test Gas...
Page 36
CONFIGURATION & FIELD CALIBRATION 2.4.3 Cal Gas Number Fig. 2.12 Calibration Gas Number 2.4.4 Procedure:...
Page 37
FIELD CALIBRATION Fig. 2.13 Calibration Screen Fig. 2.14 Apply Gas Now...
Page 49
ALARM CONFIGURATION ALARM TYPE LATCHING MUTE AUDIBLE VISUAL LED DISPLAY Yes / No Yes / No INDICATION INDICATION LEL Warning N / A N / A N / A Flash Range High Pitch Toggle LEL (Hi) (4) Flashing Tone Hi / Conc High Pitch Toggle LEL (Hi Hi)
Page 62
CONFIGURATION & FIELD CALIBRATION CAUTION: The instrument must be re-calibrated after setting the defaults. 4.1 SWITCH THE INSTRUMENT ON Fig. 4.2 Switch ON Fig. 4.3 Fig. 4.4...