SETTINGS
D I S P L A Y
Rate
Tot
SET
Volume Units G
SET
Time Units M
SET
Tot Dec Point
Format 0.0
SET
An. output Type
SET
Fast An. Output
SET
Rate For Analog
Output:
SET
SET
Rate For High
Alarm:
SET
Rate For Low
Alarm:
SET
Pulse Output
SET
Low Pass
Digital Filter:
SET
Totalizer Reset
SET
SET-UP. Use the instructions below to set up your control. Use the
to the next, and use the
This is the power-up display.
Rate/total display (shows two seconds after power-up), Press
Use the
key to choose the volume units: milliliters (mL) liters (L) or gallons (G)
Use the
key to choose the time base for rate: per second (S), per minute (M), per hour
(H), per day (D).
Use the
key to choose number of decimal places: 0, 0.0, 0.00.
Leave "Analog Output Type" unchanged unless you intend to use the 0-5 V output.
Standard setting is for 4-20 mA. Consult factory for 0-5 V output.
Leave default ("Fast Analog Output") unchanged unless you have a demanding closed-loop
application.
Standard setting is "disabled". If you enable this output, use the
time, 50 to 1000 milliseconds. This will control how rapidly the analog output tracks flow.
Use the
and
(20 mA or 5 V) occurs. On a max 10 gpm meter, 60% would be 6 gpm, for instance.
Use the
and
energize.
On a 10 gpm meter, 90% would equal 9 gpm, for instance.
Use the
and
energize. On a 10 gpm meter, 10% would equal 1 gpm, for instance.
Use the
and
selected liters, it will be pulses per liter, etc.
This setting controls the smooth/responsive trade off of the display. Increasing the time makes
the display smoother, but slower to respond to change. Use the
seconds.
If using the meter with pulsing flows, it may be necessary to increase the time.
Use the
key to switch between totalizer reset "Enabled" and "Disabled". If the reset is
enabled, every time the
zero. Settings finished.
and
keys as described to change settings within operations.
key to set the percentage of full scale at which peak analog output
key to set the percent of full scale at which the high flow alarm will
key to set the percent of full scale at which the low flow alarm will
key to set the number of pulses (0000 to 9999) per unit. If you have
key is pressed during normal operation, the totalizer will reset to
key to move from one operation
SET
to begin programming.
SET
key to select a response
to set 1, 2, 5, 10, or 20
Page 5