Follett REF20-LB Installation, Operation And Service Manual page 26

Table of Contents

Advertisement

Parameter
Default Value
Power Alarm
Yes
Relay
Power Alarm
5 minutes
Timer
Battery Level
Data Storage
Yes
Data Full Alarm
No
Data log
Sample Rate
15 minutes
Data Duration
Data Storage
Yes
Data Full Alarm
No
Track Events
Yes
P1 Datalog
No
P2 Datalog
No
P4 Datalog
Yes
P5 Datalog
No
Defrost
Manual Defrost
No
Defrost Control
Manual only
Graphing
X Axis Range
168 hours
(hrs)
Y Axis Minimum
–1 C (30 F)
Y Axis Maximum
27 C (80 F)
Graph Show
No
Alerts
P1 Graph Display
NO
P2 Graph Display
No
26
Range
Yes, No
0 to 120 minutes
Yes, No
Yes, No
0 to 360 minutes
Yes, No
Yes, No
Yes, No
Yes, No
Yes, No
Yes, No
Yes, No
Yes, No
"Power on, manual only,disable,
time of day, compressor on
time"
1 to 384 hours
–46 to 121 C (–50 to 250 F)
–46 to 121 C (–50 to 250 F)
Yes, No
Yes, No
Yes, No
Description
Defines if relay should open/close on power alarm .
Delay before sounding the power alarm .
Battery Level
Overwrite circular data?
Alarm when data memory is full?
minutes between data sampling . 0 = Off
Overwrite circular data?
Alarm when data memory is full?
Track events with log
Log P1 to event log
Log P2 to event log
Log P4 to event log
Log P5 to event log
Force the unit into a defrost
Defines tactic for initiating a defrost
Time span for x-axis
Minimum temperature shown on graph
Maximum temperature shown on graph
Determines whether alerts are displayed on the graph
Determines whether to graph probe 1
Determines whether to graph probe 2
01134592 – REF20/25 LB, PH, BB

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ref25-lbRef20-phRef25-phRef20-bbRef25-bbJ62764

Table of Contents