Savaria Ceiling Lift Track Bridge Owner's Manual page 27

Table of Contents

Advertisement

Electromagnetic Immunity (continued)
Electromagnetic Immunity - For All Equipment and Systems
The Bridge is intended for use in the electromagnetic environment specified below.
Immunity test
Radiated RF
IEC 61000-4-3
Part No. 001300-EN, 07-m01-2022
Guidance and Manufacturer's Declaration -
IEC 60601 test
level
10 V/m
80 MHz to 2.5 GHz,
1Khz, 80% AM
Mod.
Compliance
level
10 V/m
80 MHz to 2.5
GHz, 1Khz, 80%
AM Mod.
Ceiling Lift Track Bridge Owner's Manual
Electromagnetic
environment - guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the Bridge,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
3.5
d
=
P
------ -
3
3.5
80 MHz to
d
=
P
------ -
800 MHz
10
7
800 MHz to
d
=
P
----- -
2.5 GHz
10
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres.
Field strengths from fixed RF
transmitters, as determined
by en electromagnetic site
(a)
survey,
should be less than
the compliance level in each
(b)
frequency range.
Interference may occur in the
vicinity of equipment
marked with the following
symbol:
27

Advertisement

Table of Contents
loading

Table of Contents