Maico easyTymp Operation Manual page 51

Hide thumbs Also See for easyTymp:
Table of Contents

Advertisement

Calibration properties
Calibrated
transducers:
Accuracy:
Impedance calibration properties
Probe tone
Compliance
Pressure
8111254 Rev. 7.0
Probe system:
General
Reflex Frequencies:
Ipsilateral Reflex Tone
Levels:
Contralateral Reflex
Tone Levels:
Pressure
measurement:
Compliance
measurement:
Frequencies:
Level:
Distortion:
Range:
Temperature
dependence:
Pressure dependence:
Reflex sensitivity:
Temporal reflex
characteristics:
Range:
Safety limits:
Operation Manual easyTymp
Ipsilateral and Contralateral Earphone: is
integrated in the probe system.
Probe frequency transmitter and receiver and
pressure transducer is integrated in the probe
system.
Generally the instrument is made and
calibrated to be within and better than the
tolerances required in the specified standards:
3%
5 dB for 500 to 2000Hz and +5/-10 dB for
3000 to 4000Hz
5 dB for 500 to 2000Hz and +5/-10 dB for
3000 to 4000Hz
5% or 10 daPa, whichever is greater
5% or 0.1 ml, whichever is greater
226 Hz  1%, 1000 Hz  1%
85 dB SPL 1.5 dB measured in an IEC 60318-
5 acoustic coupler. The level is constant for all
volumes in the measurement range.
Max 5% THD
0.1 to 8.0 ml
-0.003 ml/C
-0.00020 ml/daPa
0.001 ml is the lowest detectable volume
change
Initial latency = 35 ms (5 ms)
Rise time = 45 ms (5 ms)
Terminal latency = 35 ms (5 ms)
Fall time = 45 ms (5 ms)
Overshoot = max. 1%
Undershoot = max 1%
- 400 to +200 daPa.
-750 daPa and +550 daPa, 50 daPa
47
02/16