4.2 INSTALLATION OF ISC40 SENSOR IN ISC40FD
Installation of the immersion fittings involves two steps:
•
Installation of the sensor
The sensor has to be mounted in the immersion fitting (see Figure 8).
•
Installation in the application
On a railing or stanchion
»
For this purpose the immersion fitting has an optional rail mounting kit /MS1. This
kit consists of a piece of guide pipe and two pipe clamps. The guide pipe is mounted
horizontally to the stanchion using one pipe clamp. The second pipe clamp connects
the guide pipe to the immersion fitting. The distance between the pipe clamps deter-
mines the distance between the stanchion and immersion fitting (see Figure 9).
With flange
»
For this purpose a stainless steel immersion fitting can be ordered as type -
SFD (DN50) or type –SFA (2"). The PVC immersion fitting has an optional depth
adjustable mounting flange /FA (with DN50 and ANSI 2" 150 lbs hole pattern) which
is illustrated in Figure 10. You must assure that the mating flange has the same hole
pattern as the supplied flange.
ISC40FD-S
Figure 8: Installation of ISC40 sensor
in immersion fitting
IM 12D07K04-01EN-P
ISC40FD-V
Figure 9: Installation on stanchion
with /MS1
/MS1