ABQ Industrial GalvanoTest 3000 Technical Reference And Operating Manual page 31

Table of Contents

Advertisement

-
After this, use a caliper to measure diameter and
length of the de-pated wire area.
Measure dia.
"d"
of
de-plated wire area
-
Use the measured wire dia. „d" and the length „l"
of the de-plated wire area to determine the actual
deplated wire area as follows:
π π π π π
F = d x
F = Area
d = Diameter of wire
π
= 3,14
l = length of de-plated area
ElektroPhysik
Insulating tape
Measure length
"l"
of
de-plate wire area
x l
-
Use ARROW-keys for setting the actual
measuring area, e.g. 5,36 mm
-
Press SET-key to show the reading as calculated
by means of the actual measuring area, e.g. : 7,4
5,36
5,36mm²
mm²
Measuring area?
SET
7,4
Sn K 5,7 * * µm
2
.
Actual
measuring area
Reading
E-27