Press START/STOP-key or SET-key for starting
measurement. At the end of the measuring
procedure, the alphanumerical display shows
again „Measuring area?".
-
Use the ARROW-keys for setting the precise de-
plating area (which has been calculated
beforehand) and confirm by pressing SET-key.
The instrument now calculates, displays and
stores the thickness value of the coating.
Note:
Depending on the type and thickness of coating, wires
≤∅ 0,3 mm may also be measured in the electrolyte
cup.
Example for measuring wires by means of the
electrolyte cup:
-
Press START-key and use ARROW-keys for
setting the measuring area according to wire
diameter and depth of immersion.
E-26
START
4,00
Measuring area? mm²
Press START again.
START
Sn K 5,7 * * µm
Remove wire after completion of measurement.
16 mm²
8 mm²
4 mm²
2 mm²
1 mm²
0,5
mm²
0,25 mm²