4. For ISE testing, select the ISE electrode being used in setup. If your electrode is not listed,

Read Type Information

There are three main read types available as shown in the setup menus.
• Auto-Read – The meter will display the measurement as it stabilizes and lock and hold the
• Timed – The meter will read and display a measurement at the set time interval that is programmed
6
ORP, RmV and mV Setup Menus
(Default values are in bold in the table below)
Electrode ID
Sample ID
Resolution
Stability
Averaging
Read Type
Alarm**
*Only for ORP and RmV modes.
** Limit and set point alarm default set to 'off'.
Cal due alarm set to 12 hours.
use the "X" electrode with the valence (+1, +2, -1, or -2) matching your electrode for
accurate results.
ISE Types
Ag+
I-
X-
CO
Cd
NH
BF
K+
+2
-4
3
CN-
Pb
Cl-
NO
+2
measurement when it is stable (The AR icon will lock on the screen.) Once the measurement is locked,
the meter will automatically export the measurement to the data log, if the data log is enabled in
the setup menu, and to a printer or computer, if a printer or computer is connected to the meter and
enabled in the setup menu. To take a new measurement, press measure (esc)/1.
by the operator. There are two options:
1.
By checking/selecting the "Interval" box, at the set time interval, the meter will automatically
export the measurement to the data log, if the data log is enabled in the setup menu, and to
a printer or computer, if a printer or computer is connected to the meter and enabled in the
setup menu. If selected, enter the time interval value in a hours:minutes:seconds format using
the numeric keypad and press f3 (select) key.
2.
If the "Interval" box is unchecked/not selected, a single timed measurement will be taken.
The meter will apply the same saving and exporting conditions as with the set time interval to
record and export data, but lock after one timed interval has expired. Press measure (esc)/1 to
reset the countdown clock and take a new reading.
Input
Off, Manual, Auto Increment
1.0, 0 . 1
Smart, Fast, Medium, Slow
Off, Automatic Smart
AutoRead, Timed, Continuous
Limit, CalDue*, Set Point
REDOX (ORP) Cl
2
2
X
Cu
-2
+2
Br-
NO
3-
x
Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module
S
ClO
O
-2
4
2
X
F-
SCN-
+
Ca
Na+
X
+2
+2