2.4 Operating Voltage Setting

(1) JP3
The operating voltage range of the IE-703002-MC is set using JP3 in the pod. When the IE-703002-MC is used
for developing application systems using the V852, V853 and V850/Sxx products keep JP3 as its factory-set
position.
Remark The JP3 settings are as follows.
Shorted
(factory-set)
Open
When the power of the target system is off or the IE-703002-MC is operated on a stand-alone basis,
the operating voltage of the IE-703002-MC differs depending on the JP3 setting. (2-3 of JP4 shorted.)
Status of JP3
Shorted
Open
Therefore, if the IE-703002-MC is operated with an option board connected while JP3 is shorted, the
option board, which should operate at 3.3 V will be damaged because 4.7 V, which is above the
absolute maximum rating, is applied. It is therefore important to pay attention to the JP3 settings.
When the IE-703002-MC is used for developing application systems using V850/Sxx products by connecting an
option board, apply the settings specified in the user's manual of the option board used.
30
CHAPTER 2 NAMES AND FUNCTIONS OF PARTS
JP3
Set when developing V852 application system.
Operating voltage range of IE-703002-MC is 3.0 to 5.5 V.
Setting prohibited when developing V852 application system.
Operating voltage range of IE-703002-MC is 3.0 to 3.6 V.
Operating Voltage of IE-703002-MC
4.7 V
3.3 V
User's Manual U11595EJ5V0UM
Description