3.2. Specifications Overview ....................4 Getting Started ........................5 Chapter 4 4.1. Unpacking and Inspection ................... 5 4.2. Setting up the CM401 ....................6 4.2.1. General Setup ........................6 4.2.2. System Installation ........................ 6 Capturing an Image ......................9 Chapter 5 Maintaining the CM401 ....................
CM401 Hyperspectral Imaging System Chapter 1: Warning Symbol Definitions Warning Symbol Definitions Chapter 1 Below is a list of warning symbols you may encounter in this manual or on your device. Symbol Description Direct Current Alternating Current Both Direct and Alternating Current...
Any modification or servicing of this system by unqualified personnel renders Thorlabs free of any liability. Only personnel authorized by Thorlabs and trained in the maintenance of this equipment should remove any covers or housing or attempt any repairs or adjustments.
Chapter 3: Description Description Chapter 3 Thorlabs’ Hyperspectral Imaging System provides a platform to study a sample with co-located features. In hyperspectral imaging, a stack of spectrally separated, two-dimensional images is acquired. This allows quick sample identification and analysis. The Hyperspectral Imaging System consists of Thorlabs’ flexible Cerna ®...
CM401 Hyperspectral Imaging System Chapter 3: Description 3.2. Specifications Overview The following table provides an overview of the CM401 specifications. For detailed specifications on individual devices, see the Specifications Chapter on page 15. Specification Value Microscope Cerna Series Modular Microscope...
This section is provided for those interested in getting the Hyperspectral Imaging System up and running quickly. 4.1. Unpacking and Inspection Open the package and carefully remove the CM401 and its accessories. The table below lists the standard accessories shipped with the device. Name...
• Acquire images using actively isolated or anti-vibrational table platforms. Failure to utilize appropriate platforms results in poor quality images. Thorlabs offers a wide array of optical tables and platforms that provide the necessary stability for your imaging needs. •...
Page 10
CM401 Hyperspectral Imaging System Chapter 4: Getting Started 3. Mount the objective, camera, and condenser onto the microscope body. Figure 4-2 Mounting the Accessories 4. Mount the rigid stand onto the optical table. 5. Mount the MCM3001 Controller Boxes (2X) onto the optical table.
Page 11
CM401 Hyperspectral Imaging System Chapter 4: Getting Started C. Cable Connections 1. Connect the MCM3001 controllers to the Controller Boxes, as shown in the figure. 2. Connect the cables form the rigid stand (X and Y axes) and the motorized objective arm to one of the MCM3001 Controller Boxes.
OK. Figure 5-1 Standard Device Requirement for CM401 3. Use the Z Control panel to focus on the sample. 4. In the Capture Setup tab, make sure that the Camera Control panel, Kurios Filter panel, and Z Control panel are visible.
Page 13
CM401 Hyperspectral Imaging System Chapter 5: Capturing an Image 5. In the Kurios Filter panel, set the value for Start, Stop, and Step Size. You can set the Start and Stop values between 420-730 nm. Figure 5-2 displays a Start value that is set to the minimum wavelength (420 nm) of the KURIOS filter.
Page 14
CM401 Hyperspectral Imaging System Chapter 5: Capturing an Image Figure 5-3 Camera Control Panel 8. Enter the Stop value as the Wavelength (for example, 730 nm). 9. In the Camera Control panel, change the Exposure value until you see the desired brightness on the image.
Page 15
CM401 Hyperspectral Imaging System Chapter 5: Capturing an Image Figure 5-5 Hyperspectral Sequence 11. Choose an arbitrary value between the Start and Stop wavelengths. In the Kurios Filter panel, enter the arbitrary value for the Wavelength. 12. In the Camera Control panel, change the Exposure value until you see the desired brightness on the image.
Page 16
CM401 Hyperspectral Imaging System Chapter 5: Capturing an Image 15. In the Capture tab, select Hyperspectral for Capture Mode. This option is only visible if you select Scientific Camera as the Image Detector and ThorKurios as the Spectrum Filter in the Hardware Connection window.
Maintaining the CM401 Chapter 6 This unit does not need regular maintenance. If you suspect a problem with your CM401, please contact our nearest office for assistance from an application engineer (see the Thorlabs Worldwide Contacts Chapter on page 24 for details).
CM401 Hyperspectral Imaging System Chapter 7: Specifications 7.8. Camera Specifications Specification Value Monochrome Item # CS126MU Sensor Type CMOS Monochrome Effective Number of Pixels (Horizontal x Vertical) 4096 (H) x 3000 (V) (~12. 3 MP) Pixel Size 3.45 µm x 3.45 µm Optical Format 1.1"...
CM401 Hyperspectral Imaging System Chapter 7: Specifications 7.9. Motorized Controller Specifications Specification Value Item # MCM3001 Motor Output Motor Drive Voltage 24 V 7.0 A (Peak) Motor Drive Current 3.0 A (RMS) Motor Drive Type 12-Bit PWM Control Control Algorithm...
CM401 Hyperspectral Imaging System Chapter 7: Specifications 7.10. Plasma Light Source Specifications Specification Value Item # HPLS343 Physical 14.23" x 7.40" x 7.87" Dimensions (L x W x H) (361.5 mm x 188.0 mm x 200.0 mm) 10 to 30 °C (Open and Closed Loop) Operating Temperature 10 to 35 °C (Eco Mode)
Waste Treatment is Your Own Responsibility If you do not return an “end of life” unit to Thorlabs, you must hand it to a company specialized in waste recovery. Do not dispose of the unit in a litter bin or at a public waste disposal site.
CM401 Hyperspectral Imaging System Chapter 10: Thorlabs Worldwide Contacts Chapter 10 Thorlabs Worldwide Contacts For technical support or sales inquiries, please visit us at www.thorlabs.com/contact for our most up-to-date contact information. USA, Canada, and South America UK and Ireland Thorlabs, Inc.
Need help?
Do you have a question about the CM401 and is the answer not in the manual?
Questions and answers