Dx-D 60C, Dx-D 60G - AGFA DX-D 60C User Manual

Hide thumbs Also See for DX-D 60C:
Table of Contents

Advertisement

DX-D 60C, DX-D 60G

Commercial name
Electrical connection DR Detector
Rated power supply
(powered by System Control Unit)
Power consumption
Environmental conditions (during normal operation)
Room temperature
Humidity (non condensing)
Atmospheric pressure
Environmental conditions (during storage and transport)
Temperature (ambient)
Humidity (non condensing)
Atmospheric pressure
Warming-up time
30 minutes
Dimensions
Dimensions
width x length x height
Weight
Maximum load
Maximum total load
Vibration tolerance
Shock tolerance
DX-D 60C, DX-D 60G | Technical Data | 79
DX-D 60C, DX-D 60G
DC +24 V, Max. 1.0 A
max. 24 W
between +10 °C and +35 °C
between 30% and 85% RH
(non-condensing)
between 700 hPa and 1060 hPa
between -15 °C and +55 °C
between 10% and 90%
(non-condensing)
between 500 and 1060 hPa
approx. 460 x 460 x 15.5 mm
4.2 kg
100 kg on an area of 40 mm in diameter
150 kg over the whole detector surface
2 G during normal operation
5G during storage and transport
20 G during normal operation
0294B EN 20190221 1131

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dx-d 60g6007/1106007/111

Table of Contents