AGFA DX-D 60C User Manual page 87

Hide thumbs Also See for DX-D 60C:
Table of Contents

Advertisement

DX-D 60C, DX-D 60G | Remarks for HF-emission and immunity | 87
Magnetic field
30 A/m
at the supply
frequency
(50/60 Hz) in
accordance
with IEC
61000-4-8
Tests of Resist-
ance to Disrup-
tion
Conducted high
frequency dis-
turbance varia-
bles in accord-
ance with
IEC 61000-4-6
Radiated high
frequency dis-
turbance varia-
bles in accord-
ance with
IEC 61000-4-3
Note: The higher value will apply at 80 MHz and 800 MHz.
Note: These Guidelines may not apply to all situations. The dis-
persion of electromagnetic waves is influenced by absorption
and reflections from buildings, objects and people.
30 A/m
IEC 60601
Level of
Test Level
Agree-
ment
3 V
150
3 V
150
eff
eff
kHz to 80
kHz to 80
MHz
MHz
6 V
in
6 V
in
eff
eff
the ISM
the ISM
bands 150
bands 150
KHz to 80
KHz to 80
MHz
MHz
3 V/m
3 V/m
80 MHz to
80 MHz
2.7 GHz
to 2.7
GHz
Magnetic field at the net-
work frequency should
correspond to the typical
values as they are in a
commercial and clinical
environment.
Electromagnetic Environment
The electromagnetic field
strength of a stationary RF
transmitter determined by an
electromagnetic test survey
must be less than the compli-
ance level of each frequency
range.
Interference may occur in the vi-
cinity of equipment marked
with the symbol:
0294B EN 20190221 1131

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dx-d 60g6007/1106007/111

Table of Contents