INSTALLING THE IN-FASTENER SHUT-OFF

English
Note: These installation instructions are intended for cots that
you will NOT use with Power-LOAD. For Model 6516 cots with
the Power-LOAD option, see the Power-LOAD Operations/
Maintenance Manual for installation instructions.
WARNING
The in-fastener shut-off must be positioned properly before
placing the cot into service. Failure to install the in-fastener
shut-off may cause injury to the patient or operator and/or
damage to the vehicle.
The cot and fastener system have an integrated in-fastener
shut-off function that disables the cot motor when the cot is
secured into the cot fastener. Securely tighten the bolts on
the fastener before installing the shut-off bracket. Install the
shut-off bracket onto the rail clamp assembly before putting
the cot into service.
1.
Place the cot into a loading position (any position where
the load wheels of the head section meet the vehicle
floor height).
2.
Lift the vehicle bumper to the raised position (if equipped).
3.
Roll the cot to the open door of the patient compartment.
4.
Push the cot forward until the load wheels are on the
compartment floor and the safety bar passes the safety
hook.
3.
For maximum clearance to lift the base, pull the cot back
until the safety bar engages the safety hook.
4.
Raise the base and push the cot into the patient
compartment
instructions.
5.
Engage the extended head section of the cot into the cot
fastener antler and secure the cot post into the fastener
rail clamp.
6.
Adjust the shut-off bracket along the rail clamp until the
"diamond" (A) on the sensor housing is lined up with the
pop rivet head (B) as shown in Figure 10.
7.
Using a T27 Torx driver, securely fasten the bolts to
attach the shut-off bracket to the rail clamp assembly.
8.
Press the retract (–) button to ensure that the motor does
not turn on while the cot is in the fastener. The battery
indictor will still illuminate. If the motor turns on, readjust
the shut-off bracket.
Return To Table of Contents
1-22
Cot Fastener Installation
following
the
appropriate
Note: Align the 'diamond' (A) on the sensor
housing cover with the pop rivet head (B) on the
in-ambulance shut-off.
Figure 10: Cot Engaging Cot Fastener
loading
Figure 11: In-Fastener Shut-Off Module
WARNING
• Do not attempt to operate the cot when it is loaded
into a cot fastener.
• The in-fastener shut-off is only a means for
disabling the electronic functionality. Damage to
the product and/or injury to the patient and/or
operator may occur if used for any other purpose.
• Any emergency vehicle to be used with this cot
must have the in-fastener shut-off system installed
(if not using Power-LOAD).
6516-109-005 REV B
B
A
www.stryker.com

Advertisement

Table of Contents
loading

Table of Contents