WARNING
Using this unit for medical purposes
The connectors on this unit are not isolated.
Do not connect any device other than one which conforms
to IEC 60601-1 standards.
When an information technology device or AV device that
uses an alternating current is connected, current leakage
may result in an electric shock to the patient or operator.
If use of such a device is unavoidable, isolate its power
supply by connecting an isolation transformer, or by
connecting an isolator between the connecting cables.
After implementing these measures, confirm that the
reduced risk now conforms to IEC 60601-1 standards.
Caution
This unit with the connector cover attached conforms to
the waterproofed standard. (LMD-X310MT: See page 15,
LMD-X550MT: See page 16)
Be sure not to use the unit with the connector cover
removed since waterproofed performance is not
guaranteed.
LMD-X310MT only
Caution
Connect the DC connector to the unit, then connect the
power cord of the AC adaptor.
To remove the DC connector, remove the power cord of
the AC adaptor, then remove the DC connector.
WARNING
For the DC power supply, make sure to use the supplied
AC adapter, AC-300MD.
If another power supply is used, there is a risk of fire or
electric shock.
LMD-X550MT only
WARNING
To avoid the risk of electric shock, this
equipment must only be connected to a supply
mains with protective earth.
8
GB
WARNING
This unit has no power switch.
To disconnect the main power, unplug the power plug.
When installing the unit, incorporate a readily accessible
disconnect device in the fixed wiring, or connect the
power plug to an easily accessible socket-outlet near the
unit.
Do not position the ME equipment where it is difficult to
unplug the power plug.
If a fault should occur during operation of the unit,
operate the disconnect device to switch the power supply
off, or disconnect the power plug.
Warning on power connections
Use a proper power cord for your local power supply.
1. Use the approved Power Cord (3-core mains lead) /
Appliance Connector / Plug with earthing-contacts
that conforms to the safety regulations of each country
if applicable.
2. Use the Power Cord (3-core mains lead) / Appliance
Connector / Plug conforming to the proper ratings
(Voltage, Ampere).
If you have questions on the use of the above Power Cord /
Appliance Connector / Plug, please consult a qualified
service personnel.
Warning on power connections for
medical use
Customers in the U.S.A. and Canada should use the
following type of power cord.
Customers in other countries or regions should use the
power cord prescribed by their country or region.
Plug type
Cord type
Minimum rating for plug
and appliance couplers
Safety approval
* Note: Grounding reliability can only be achieved when
the equipment is connected to an equivalent receptacle
marked "Hospital Only" or "Hospital Grade".
U.S.A. and Canada
*
HOSPITAL GRADE
Min. Type SJT
Min. 18 AWG
10 A / 125 V
UL Listed and CSA