Features; Full Feature Test - Arjohuntleigh Maxi Sky 2 Maintenance And Repair Manual

Hide thumbs Also See for Maxi Sky 2:
Table of Contents

Advertisement

4 - 4
INTRODUCTION
This document covers the maintenance and repair of the Maxi Sky 2 and Maxi Sky 2 Plus.
Maxi Sky 2 has a maximum capacity of:
- 272 kg (600 lb).
The Maxi Sky 2 Plus has a maximum capacity of:
- 454 kg (1000 lb) when set in dual mode.
- 272 kg (600) when set in single mode.
CONFIGURATION MAXIMUM SWL REFERS TO THAT OF THE LOWEST RATED

FEATURES

The emergency features of the Maxi Sky 2 ceiling lift include:
- Stop.
- Overload current limit protection in the PCB.
- Upper limit switch.
- Lower limit switch/no tension limit.

FULL FEATURE TEST

These tests will be performed after replacement of a part or a repair is performed to ensure the quality of the repair. The
test will be called out throughout section 6.
#
Function
Item
1
Charging
2
Horizontal
displacement
3
Vertical
displacement
4
Inclined PDPS
5
High / Low limit
switch
6
Hand control
7
Up / Down
To perform a Full Featur Test on the MS2 Plus, refer to Maxi Sky 2 Plus document "Initial Start Up" 001-31229-EN section
14 "Testing the Maxi Sky 2 Plus System".
THE SAFE WORKING LOAD (SWL) WILL DEPEND ON THE LIFT'S
ATTACHMENTS, AS WELL AS RAIL INSTALLATION
Activation/
Validation
Manual or button Push R/L button or direct lift
manually on a charging station
Buttons
Push right and left function on
the hand control
Buttons
Push up and down function on
the hand control
Buttons
Push inclined or sitting position
Buttons
Push up and down function on
the hand control
Buttons
Press each button
Buttons
Press alternate buttons on lift
NOTE
- Control panel on the unit / LED display
- Manual emergency lowering mechanism.
- Audible variable low-battery warning.
- Malfunction & maintenance code.
Actions
Criteria for approval
Activation of the charging station
For ECS option, system shall charge when
battery level is under 80%
No abnormal noise. Lift is moving along the rail
in the expected direction.
No abnormal noise. Spreader bar is moving up
and down.
No abnormal noise. Powered DPS is moving
according to the wanted action.
High limit switch must activate at top of stroke.
No abnormal noise. Low limit switch must
activate at end of stroke.
No tension limit must stop lowering action if
there is no tension on strap (e.g. without ac-
cessories)
All functions are working: no display issues.
Strap is moving up and down