QUICK REFERENCE GUIDE - CLEANING AND DISINFECTION
Maxi Sky 2
WARNING
To avoid eye and skin damage, always use protective glasses and gloves. If contact
occurs rinse with plenty of water. If eyes or skin becomes irritated, contact a physician.
Always read the instructions for use and material safety data sheet of the disinfectant.

Allowed disinfectant agents

Basic Ceiling Lifter model and Maxi Sky 2
Ceiling Lifter model (woven nylon strap)
Maxi Sky 2 Infection Control model
(non-porous strap)
Other Accessories
* and **
Medium 2-hook spreader bar
(700-05461-BOX) & 4-hook spreader bar
(700-05443)
Stretcher Frame (700-19522) & Morgue
stretchers (700.05380, 700.05385,
700-05382 & 700-05387)
Manual Open DPS (700-19340-BOX)
Medium 2-hook spreader bar
(700-19415)
Small 2-hook spreader bar
(700-19420) & Medium 2-hook IC
spreader bar (700-19470)
4-hook spreader bar (700-19430) &
Stretcher Frame (700-15695)
Manual Flat DPS
(700-19475 & 700-19200)
Manual Open DPS (700-19480)
Medium PDPS (700-19350) &
Large PDPS (700-19355)
* Metal portion of the WCS connector may corrode after repeated exposure to bleach & hydrogen peroxide.
** Handset, WCS and PDPS cables may become slightly grimy after repeated contact with bleach or hydrogen
peroxide.
*** The cover for this particular model may be vinyl padding or plastic. If vinyl, do not use these disinfecting agents.
Quaternary
Isopropyl
Ammonium
Alcohol
≤ 0.21%
≤ 70%
(2,100 ppm)
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
001-15796 Rev 0 • 05/2020 • Page 1/2
Phenolic
Bleach
≤ 1.56%
≤ 1%
(15,600 ppm)
(10,000 ppm)
X
X
X
X
X
X
X
X
X
X
X***
X
X
X
X
X
X
X
X
X
X
Peracetic
Hydrogen
Acid
Peroxide
≤ 0.25%
≤ 1.5%
(2,500 ppm)
X
X
X
X
X***
X***
X
X
X
X
X
X
X
X
X
X

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Maxi Sky 2 and is the answer not in the manual?

Questions and answers

Summary of Contents for Arjohuntleigh Maxi Sky 2

  • Page 1 ≤ 70% (15,600 ppm) (10,000 ppm) ≤ 1.5% (2,500 ppm) (2,100 ppm) Basic Ceiling Lifter model and Maxi Sky 2 Ceiling Lifter model (woven nylon strap) Maxi Sky 2 Infection Control model (non-porous strap) Other Accessories * and ** Medium 2-hook spreader bar (700-05461-BOX) &...
  • Page 2 • Strap Disinfect Wipe the Maxi Sky 2 and accessories using a solution containing one of the compatible disinfectants. Disinfection methods should comply with the local or national guidelines, depending on the Healthcare Facility or country of use. If uncertain, consult local Infection Control specialist.

Table of Contents