the result stabilization.
5. When there is no weighing, but the balance should be ready for operation, it may be
switched
off by I/
key. It causes the balance reading system deactivation end entering the
standby mode. Balance turning on is performed by pressing "I/ " key.
6. Protect the balance from dust, aggressive dusts and liquids. For cleaning purposes
use water and dry it.
10. Balance test
During balance operation, in order to confirm its efficiency, it is recommended to check
the weighing precision by putting and object of exactly known weight before and after
series of measurements.
For testing of verified balances use weight standard, having valid standardization
certificates. In case of allowable measurement error exceeding contact authorized
service company to perform balance adjustment.
11. Balance adjustment
Balance adjustment must be performed by authorized service company
only, as it is connected with necessity of seals breaching, required for
warranty purposes.