Data exchange method:
Balance indication reading (corresponds to
Computer Balance: S I CR LF (53h 49h 0Dh 0Ah) – initiation signal. Balance Computer:
balance sends 16 bytes of data according to description:
Byte
1
- sign "-" or space
Byte
2
- space
Byte
3 4 - digit or space
Byte
5÷9 - digit, comma or space
Byte
10
- digit
Byte
11
- space
Byte
12
- k, l, c, p or space
Byte
13
- g, b, t, c or %
Byte
14
- space
Byte
15
- CR
Byte
16
- LF
Tare setting from computer (corresponds to T key use):
Computer Balance: S T CR LF (53h 54h 0Dh 0Ah),
Balance Computer: no reply.
 B alance zeroing (corresponds to 0 key use):
Computer Balance: S Z CR LF (53h 5Ah 0Dh 0Ah),
Balance Computer: no reply.
Balance switching on/off (corresponds to I/ key
Computer Balance: S S CR LF (53h 53h 0Dh 0Ah),
Balance Computer: no reply.
 M enu displaying (corresponds to MENU key use):
Computer Balance: S F CR LF (53h 46h 0Dh 0Ah),
Balance Computer: no reply.
 S etting of threshold 1 value (option):
Computer Balance: S L D1...DN CR LF (53h 4Ch D1...DN 0Dh 0Ah) where: D1...DN –
threshold value, maximum 8 characters.
Balance Computer: no reply.
Example:
To set 1000g in balance B1.5 (d=0.5g) enter:
S L 1 0 0 0 . 0 CR LF (53h 4Ch 31h 30h 30h 30h 2Eh 30h 0Dh 0Ah). To set
100kg in balance B150 (d=50g) enter:
S L 1 0 0 . 0 0 CR LF (53h 4Ch 31h 30h 30h 2Eh 30h 30h 0Dh 0Ah),
Setting of threshold 2 value (option):
Computer Balance: S H D1...DN CR LF (53h 48h D1...DN 0Dh 0Ah), where:
D1...DN – threshold value, maximum 8 characters.
Balance Computer: no reply.