Calibration of the balance

If the analytical balance PCE-ABT 220L is inaccurate, it must be calibrated.
Follow these steps:
Position the balance on a stable, vibration-free surface. In addition, the balance must
remain stationary for one hour to allow it to adjust to the ambient temperature.
Press the CAL key. "CAL-200" is displayed. Then place the 200 g weight on top. Initially,
the balance displays "-------". The balance then shows 200.0000 g. This means that the
calibration is finished.
8

Settings

8.1

Sensitivity

With the power on, press ON until the word "Stab" appears on the display. Press TARE to
access the setting options; press the ON key to change the setting:
Lowest sensitivity
Low sensitivity
Normal sensitivity
High sensitivity
Press TARE to save the setting.