Transport And Storage Conditions - AGFA Classic E.O.S. Installation Planning

Hide thumbs Also See for Classic E.O.S.:
Table of Contents

Advertisement

Installation Planning
9.3
Chapter 14 / 34
Radiation effects
Floor conditions
Floor load
Cleaning sink
Heat emission
(approx. values)

Transport and storage conditions

Ambient
temperature
Classic E.O.S. / Classic E.O.S. CL
(Type 5270 / 100 / 105)
Upon machine installation in the close vicinity to
the X-ray room, the local regulations for radiation
protection must be followed (protection of
personnel against scattered radiation). The
machine has not been designed for the
installation in the X-ray room, i.e. there is no
internal screening against scattered radiation.
Waterproof, chemical-resistant floor covering
(pH value 4 - 11)
A floor drain close to the film processor is
recommended.
50 N / cm² (7.75 N / in²)
A cleaning sink with water tap and hose shower
should be provided for maintenance work.
Minimum inside dimensions of the sink:
Width
70 cm (27.56 inch)
Depth
40 cm (15.75 inch)
Height
20 cm (7.87 inch)
Standby
into the room
(max.)
Film run
into the connected
(max.)
exhaust
into the room
total
Exhaust
Integrated exhaust in the dryer is a
connect.
standard feature.
During film cycle the exhaust runs at full
capacity. During standby mode the
exhaust can be set to half-capacity via
the service program.
up to max. -25 °C for 72 h
max. 55 °C for 96 h
Relative humidity in the given temperature range
between 5% and 95%.
DD+DIS303.03E
250 W / 900 kJ/h
1200 W / 4320 kJ/h
900 W / 3240 kJ/h
2100 W / 7560 kJ/h
(IEC 68-2-1 (Ab))
(IEC 68-2-2 (Bb))
Edition 6, Revision 0