R&S Scope Rider RTH Digital Oscilloscope
1 Revision History
Date
29.06.2016
03.05.2016
17.02.2016
07.01.2016
12.11.2015
2 Firmware Update
The firmware update file for the R&S RTH is a single rsi-file.
2.1 Preparing the Installation
Please transfer the following file to a USB memory stick:
RTH_Update_1.30.3.5.rsi
2.2 Updating the Firmware on the Instrument
Follow these steps to update the instrument firmware:
1. Make sure that the instrument has sufficient battery power or connect the
instrument to the mains supply.
2. Insert the memory stick into the USB port of the oscilloscope
3. Press the
4. Scroll down to "Firmware Update" > "Open File"
5. In the path menu navigate to the location of the RTH_Update_xxx.rsi file on your
memory stick and press "Select"
6. Confirm with "Yes" the start of the update process.
7. Wait until the update process is finished. This may take up to 15 minutes. The
instrument automatically reboots itself once the update process is finished.
IMPORTANT INFORMATION for update from version V0.90.8.30 to any other version:
For update from V0.90.8.30 to this version the instrument requires a second and
manual boot once the update is finished. The instrument cannot complete the first
automatic boot at the end of the update and will freeze displaying the bootscreen.
Please remove the battery, insert the battery again and then switch on the instrument.
After this action the instrument will boot correctly.
Interruption of the power during the update process can make the instrument unusable!
Rel. Note rev.
Changes
5.0
Firmware Release Version V1.30.3.5 for RTH1002 and
RTH1004
4.0
Firmware Release Version V1.20.3.4 for RTH1002 and
RTH1004
3.0
Firmware Release Version V1.10.2.10 for RTH1002 and
RTH1004
2.0
Firmware Release Version V1.0.1.19 for RTH1002 and
RTH1004
1.0
First version RTH1004 firmware V0.90.8.30
Firmware image for R&S
key or tap on the "Menu" button and select "Setup"
®
®
RTH1004 and R&S
RTH1002