6 Known Issues

The following table provides a more detailed overview about known issues. It also
indicates the version in which the issue appeared for the first time.
Version
1.30.3.5
Release Notes 1326.2645.00 - 5
Function
Solved: Option K1, K2 threshold setting: It is necessary to first select the channel and only
after that set the threshold. If the threshold is entered before selecting a new channel, the
threshold value will not be applied.
Solved: For time bases between 500 s/DIV ... 50 ms/DIV the instrument seems very slow as it
only starts to display the trace after the acquisition has been fully completed and processed
by firmware.
Options K1, K2 and K19: several problems solved with triggering, also in combinations of K1
or K2 with K19
Solved: Option B1: Measurements on digital channels are not supported.
Solved: K15, History player cannot be stopped in mode automatic with 5000 segments
Solved: Option B1: In stop mode artefacts may be visible on digital channels.
Some issues with the autoset functionality were solved.
Solved: RTH1002, indication of statistics in frequency measurement (min, max, average over
all frequencies) may not be correct in some cases
Solved: RTH1002/RTH1004 full selfalignment. In rare cases the AB alignment step fails. In
this case repeat the full alignment. Typically the second try will pass.
Solved: Very rarely, if instrument is switched on and idle for a certain time: instrument reacts
very slowly or not at all to keypad and touchscreen operation and needs to be rebooted.
Solved: Using Cursor with Math as input channel in combination with zoom is not supported.
Combination can be selected, but leads to incorrect behavior of the instrument.
Option B1, assignment of channel indicators corrected; user can now move digital channels
on the screen
Support added on RTH1002: temperature measurement using PT500 platinum sensor
Improvement in trigger hysteresis, especially for 10:1 probe settings.
Solved: Option K2, UART: the value of the parity bit is not correct
Solved: Option K1, I2C bus triggers: trigger types address, data and address and data do not
trigger reliably
Solved: Very sporadic incorrect trigger on the analog channels
Solved: The battery controller consumes very much energy from the battery when the RTH is
switched off and disconnected from the mains. Therefore the battery runs empty very quickly.
Solved: Screenshot button: if pressed a second time before the pop-up window disappears
the instrument will freeze.
Solved: Logger does not provide a backup of the log results yet if the RTH is switched off
during logging.