Placing your network
receiver for best
performance
Activating your network
Thermo Fisher Scientific
9. Click on Connection in the left-hand panel to select the local TCP
communication port for your receiver. Each receiver must have a
unique local TCP port number.
10. In the Endpoint configuration zone, enter the desired TCP port
number in the Local Port field (e.g., 10001 for your first receiver,
10002 for the second, etc.). Click on OK.
11. Click on Apply Settings in the left-hand panel to save your
configuration changes to the network receiver. You may close your
browser after the configuration is saved.
12. Re-start the receiver by unplugging the AC adapter and plugging it
back in.
For optimal operation, follow these placement recommendations:
•
To avoid electrical interference, place the receiver at least 80 cm (31 in.)
away from electrical outlets, wiring and network cabling.
•
For best signal propagation, place the receiver in an open area at a
slightly elevated location (i.e., not directly on the ground), about 2
meters (6.5 ft.) high.
The Smart-Vue Network Receiver is a standalone device that requires mains
power for operation. To activate your receiver:
receiver
1. Screw the provided antenna onto the network receiver without
forcing it.
2. Place the receiver as described in the previous section, and use the
wall-mount adapter or Velcro® tape to adhere it.
3. Plug the AC adapter into the electrical power outlet or UPS.
Note We highly recommend that you use a UPS for your receiver(s).
Installing drivers and port redirection software
Smart-Vue Network Receiver (TCP/IP) Installation Guide
13