ZEISS
4.2
Setting the illumination
•
To switch-on and operate the cold light source, observe the separate user manual of the light
source.
•
Switch off the light source, when not in use.
To illuminate an object proceed as follows:
•
Set a medium brightness level and adjust the stereo microscope to the object (considering
Sections 4.3 to 4.8).
•
Adjust the light spot emitted by the light guide to the object.
•
Set minimum zoom position and correctly focus the object (using the focus knob) while observing it
through the eyepieces.
•
Increase zoom magnification as demanded. Eventually refocus the object.
•
Adjust brightness of the cold light source (while observing the object through the eyepieces) until
demanded brightness level is reached.
4.3
Adjusting the Floor Stand S
Parts of the stand are located at head height, with a corresponding risk of injury. Put the device
in the transport position immediately after use (see Fig. 7) and observe sufficient headroom.
Floor Stand S allows to variably adjust the used stereo microscope to large-size or voluminous objects.
The following adjustments are possible:
− Vertical adjustment on articulated suspension arm, see Section 4.4
− Longitudinal adjustment on the swivel joints, see Section 4.5
− Tilting at the tilting head, see Section 4.6
− Rotation of the Stemi mount, see Section 4.7
4.4
Vertical adjustment on the articulated suspension arm
• Swivel the articulated suspension arm (Fig. 3/2) upwards or downwards to vary the vertical position
of the microscopy system.
The adjustment range is ±300 mm. Vertical adjustment also changes the longitudinal adjustment.
16
OPERATION
Setting the illumination
435413-7144-010
Floor Stand S