AGFA CR 35-X User Manual page 91

Table of Contents

Advertisement

This device is intended for operation in the electromagnetic
environment given below. The user of the device should ensure
that it is used in such an environment.
Resistance to
Jamming Test
Discharge of static
electricity in
accordance with
IEC 61000-4-2
Fast transient
electrical
disturbance
variables / bursts in
accordance with
IEC 61000-4-4
Impulse voltages
(surges) in
accordance with
IEC 61000-4-5
Voltage
breakthroughs,
short term
interruptions and
variations in the
voltage supplied in
accordance with
IEC 61000-4-11
Magnetic field at
the supply
frequency (50/60
Hz) in accordance
with IEC 61000-4-8
• REMARK : U
is the alternating current in the network before the application of the test level.
r
4454C EN 2013-05-14
IEC 60601
Test Level
+ 6 kV contact
+ 6 kV contact
discharge
discharge
+ 8 kV air discharge
+ 8 kV air discharge
+ 2 kV for network
+ 2 kV for network
leads
leads
+ 1 kV for entry and
+ 1 kV for entry and
outlet leads
outlet leads
+ 1 kV push-pull
+ 1 kV push-pull
voltage
voltage
+ 2 kV common
+ 2 kV common
mode voltage
mode voltage
• < 5% U
(> 95%
• < 5% U
r
breakthrough of
U
) for ½ period
r
• 40% U
(> 60%
• 40% U
r
breakthrough of
U
) for 5 periods
r
• 70% U
(30%
• 70% U
r
breakthrough of
U
) for 25 periods
r
• < 5% U
(95%
• < 5% U
r
breakthrough of
U
) for 5 s
r
3 A/m
3 A/m
Level of
Electromagnetic Environment
Agreement
Floors should consist of wood,
concrete or ceramic tiles. The
relative humidity must be at least
30%, if the floor is made of
synthetic material.
The quality of the voltage
supplied should correspond to a
typical commercial or clinical
environment.
The quality of the voltage
supplied should correspond to
that of a typical commercial or
clinical environment.
(> 95%
r
breakthrough of
The quality of the voltage supply
U
) for ½ period
r
should correspond to that of a
(> 60%
r
typical commercial or clinical
breakthrough of
environment.
U
) for 5 periods
If the user wants the device to
r
work continuously, even when
(30%
r
the energy supply is interrupted,
breakthrough of
it is recommended to use an
U
) for 25 periods
r
energy supply free of
(95%
r
interruptions or a battery.
breakthrough of
U
) for 5 s
r
Magnetic field at the network
frequency should correspond to
the typical values as they are in a
commercial and clinical
environment.
Remarks for HF-emission and immunity
CR 35-X D
IGITIZER
Guidelines
91