Checking 525B Status - Fluke 525B User Manual

Temperature/pressure calibrator
Hide thumbs Also See for 525B:
Table of Contents

Advertisement

Data Type
Integer
Integers for some controllers or computers are decimal numbers in the range
-32768 to 32768.
Responses in this range are labeled Integer.
Example:
Returns:
Floating
Numbers that may have up to 15 significant figures plus an exponent that may
range from ±E20.
Example:
Returns:
Character
This type of response is always a keyword.
Response
Example:
Data (CRD)
Returns:
Indefinite
Any ASCII characters followed by EOM. Queries with this type of response MUST
ASCII (IAD)
be the last Query in a program message.
Example:
Returns:

Checking 525B Status

You have access to status registers, enable registers, and queues in the Calibrator
to indicate various conditions in the instrument as shown in Figure 4-5. Some
registers and queues are defined by the IEEE-488.2 standard. The rest are specific
to the Calibrator. In addition to the status registers, the Service Request (SRQ)
control line, and a 16-element buffer called the Error Queue provide status
information. Table 4-11 lists the status registers and gives the read/write
commands and associated mask registers.
Status Register
Serial Poll Status Byte (STB)
Service Request Enable Register (SRE)
Event Status Register (ESR)
Event Status Enable Register (ESE)
Table 4-10. Response Data Types
Description
*ESE 123; *ESE?
123
OUT?
1.429300E+00
OUT 1V; FUNC?
DCV
*IDN?
FLUKE,525B,<serial number>,<firmware version>
Table 4-11. Status Register Summary
Remote Operation
Checking 525B Status
Read
Write
Command
Command
*STB?
*SRE?
*SRE
*ESR?
*ESE?
*ESE
4
4-21