Changing Between Remote And Local Operation; Local State; Local With Lockout State; Remote State - Fluke 525B User Manual

Temperature/pressure calibrator
Hide thumbs Also See for 525B:
Table of Contents

Advertisement

525B
Users Manual
To test the IEE-488 port:
1. Open the MET/CAL Editor application.
2. Select File->New or press Ctrl+N.
3. Enter IEEE [@4]*IDN?[I$]. If you did not select address 4 in step 6 above,
substitute the address you entered for the 4.
4. Select Edit->Name Procedure or press Ctrl+I and enter Test 525B GPIB.
5. Select Compile->Next Error or press F9.
6. Select Test Run->Restart or press F2.
7. Check the S-Reg check box.
8. Select Run.
9. Verify that the S-Reg (MEM2) window displays Fluke,525B,0,<firmware
rev.>.
10. Select Quit.
11. Select File->Save or press Ctrl+S to save the procedure.
12. Exit the MET/CAL Editor.

Changing Between Remote and Local Operation

In addition to Local mode (front panel operation) and remote, the Calibrator can be
placed in a local lockout condition at any time by command of the controller.
Combined, the local, remote, and lockout conditions yield four possible operating
states described as follows.

Local State

The Calibrator responds to local and remote commands. This is normal front panel
operation. All remote commands are allowed to execute.

Local with Lockout State

Local with lockout is identical to local, except the Calibrator will go into the
remote with lockout state instead of the remote state when it receives a remote
command. You can only enter the Local with Lockout State by sending the IEEE-
488 GTL (Go to Local) message when the 525B is in remote with lockout.

Remote State

When the Calibrator is placed in remote, either via RS-232 REMOTE command,
or via IEEE-488 asserting the REN line, it enters the remote state. The top line of
the display changes to: REM.