Range
Resolution
Accuracy
±(0.8% of rdg+3 digits)
200Ω
0.1Ω
2kΩ
20kΩ
10Ω
±(0.8% of rdg+1 digit)
200kΩ
100Ω
2MΩ
1kΩ
±(1.0% of rdg+2 digits)
20MΩ
10kΩ
±[5.0% of (rdg-10digits) +10 digits]
200MΩ
0.1MΩ
Overload protection: 380V dc or 380Vac rms.
Open Circuit Voltage: Less than 700mV.
4.2.5 Diode
Range
Resolution
Function
Display read approx. forward
1mV
voltage of diode
Forward DC Current: approx. 1mA
Reversed DC Voltage: approx. 2.8V
Overload protection: 380Vdc or 380Vac rms.
4.2.6 Audible continuity
Range
Continuity beeper
≤30Ω
Open circuit voltage: Less than 700mV.
Overload protection: 380Vdc or 380Vac rms.

4.2.7 Transistor

Range
Description
Test Condition
Display read approx.
Base Current approx.
10 μ A, Vce approx.
HFE value (0-1000) of
hFE
transistor under test (all
2.8V.
type).

4.2.8 Temperature

Range
Resolution
±(5.0% of rdg+4 digits)
-20℃~0℃
±(1.0% of rdg+3 digits)
1℃
1℃~400℃
±2.0% of rdg
401℃~1000℃
20
Overload protection: Resettable fuse (F200mA/250V).
* Temperature specifications do not include thermocouple
errors.

4.2.9 Capacitance

Overload protection: Resettable fuse (F200mA/250V).

4.2.10 DC Current

Overload protection: Resettable fuse (F200mA/250V). 10A
Max. input current: 200mA dc or 200mA ac rms for mA range,
For measurements>5A, 4 minutes maximum ON to measure
10 minutes OFF; Above 10A unspecified.

4.2.11 AC Current

Overload protection: Resettable fuse (F200mA/250V). 10A
Max. input current: 200mA dc or 200mA ac rms for mA range,
Accuracy
For measurements>5A, 4 minutes maximum ON to measure
10 minutes OFF; Above 10A unspecified.
Frequency Range: 40Hz-400Hz
Response: Average, calibrated in rms of sine wave

5. MAINTENANCE

Range
Resolution
Accuracy
2nF
1pF
20nF
10pF
±(4.0% of rdg+3 digits)
200nF
0.1nF
2μF
1nF
20μF
10nF
Range
Resolution
Accuracy
20mA
10μA
±(1.5% of rdg+1 digit)
200mA
0.1mA
±(2.0% of rdg+5 digits)
10A
10mA
range unfused
10A dc or 10A ac rms for 10A ranges.
Range
Resolution
Accuracy
±(1.0% of rdg+3 digits)
2mA
1μA
±(1.8% of rdg+3 digits)
200mA
0.1mA
±(3.0% of rdg+7 digits)
10A
10mA
range unfused.
10A dc or 10A ac rms for 10A ranges.
21