8.1 Technical Specifications

All specifications herein mentioned apply to the DS2002A series oscilloscopes. Before checking
an oscilloscope from AD INSTRUMENTS to see if it complies with these specifications, make sure
it meets the following conditions:
The oscilloscope must have been operating continuously for twenty minutes under the
specified operating temperature.
The Do Self Cal operation must be performed through the Utility menu if the operating
temperature changes by more than 5℃.
The oscilloscope must be within the factory calibration interval.
All specifications are guaranteed unless noted 'typical'.
Oscilloscope Specifications
Horizontal
Sample Rate Range
Waveform Interpolation
Record Length
SEC/DIV Range
Sample Rate and
Delay Time Accuracy
Delta Time Measurement
Accuracy
(Full Bandwidth)
Position Range
DS2002A Series Digital Storage Oscilloscope User Manual

Chapter 8 Specifications

500MS/s--1GS/s
(sin x)/x
Maximum 1M samples per single-channel; maximum 512K
samples per dual-channel (4K,16K,40K optional)
DS2062A
DS2102A
DS2102A
4ns/div to 40s/div, in a 2, 4, 8
sequence
±900ppm over any ≥1ms time interval
Single-shot, Normal mode
± (1 sample interval +100ppm × reading + 0.6ns)
>16 averages
± (1 sample interval + 100ppm × reading + 0.4ns)
Sample interval = s/div ÷ 200
DS2062A DS2102A
DS2102M
4ns/div to 8ns/div
20ns/div to 80μs/div
200μs/div to 40s/div
DS2202A
2As/div to 10ns/div
Specifications
DS2202A
2As/div to 40s/div, in a 2, 4, 8
sequence
(-8div × s/div) to 20ms
(-8div × s/div) to 40ms
(-8div × s/div) to 400s
(-4div × s/div) to 20ms
62