Specifications - ADInstruments AD511 Instrument Manual

Professional insulation tester
Table of Contents

Advertisement

Available languages

Available languages

Specifications

Safety and Compliances
Certification
Compliances
Physical Specifications
Display (LCD)
Digital: 9999 counts
Analog bar graph.
Operating Temperature
0 ~40
Storage Temperature
-20 ~60
Relative Humidity
85%
90%
Battery Type
8pcs of 1.5V (LR14) batteries or DC15V power adaptor.
DC15V power adaptor is optionally at extra cost.
Dimensions (H x W x L)
202 x 155 x 94 mm
Weight
Approx. 2kg (including battery)
General Specifications
Range
Auto
Display OL on insulation resistance range
Overloading
Battery Indicator
Display
Icon Display
Equips with function and battery indicator icons.
Current Consumption
Maximum: around 90mA
Average: around 20mA
Feature Summary
Display Backlight
Autorange
Warning
Voltage
COMP Measurement
PI Measurement
Detailed Accuracy Specifications
Accuracy:
([% of reading] + [number of least significant digits), guarantee for 1 year.
Operating temperature: 18 ~28
Relative humidity: 45~75%RH
A.
Voltage Measurement
Measurement Range
Resolution
Accuracy
B.
Insulation Resistance Measurement
Output Voltage
100V
Display Range
1M:a99.9M:
100~500M:
Open Circuit Voltage
DC1000V + 20%, -0%
Test Current
1mA~1.2mA @100k:
Short Circuit
Accuracy
Caution
At any output voltage, when the tested resistance is les than 5M:, the testing time cannot exceed 10 seconds.
C.
Low Resistance Measurement
Function
Measurement Range
Resolution
Accuracy
Maximum open circuit voltage
Buzzer
Overload Protection
©Copyright Abacanto Digital SA.
_______________________________________________________________________________
IEC 61010 CAT.II 1000V, CAT.III 600V
overvoltage and double insulation
standard
(32 ~104 )
0 ~40
below;
-20 ~60 :
,
,
,
Bright backlight for clear readings in poorly lighted areas.
The Meter automatically selects best range
and red light will on.
Auto release voltage
Use the Compare function to set a pass/fail compare level
for the insulation measurements.
Polarization Index is the ratio of insulation resistance. You
can pre-set two point of times and automatically carry out the
measurement.
DC Voltage
30 ~
1000V
250V
5M:a99.9M:
100~999M:
1.00~1.99G:
DC250V + 20%, -0%
1mA~1.2mA @250k:
Manual de usuario AD511
1V
(2%+3)
500V
1M:a99.9M:
100~999M:
1.00~3.99G:
DC 500V + 20%, -0%
1mA~1.2mA @500k:
Around 2.0mA
100k: to 100M: :
(3%+5)
100M: above:
(5%+5)
Resistance
0.1:a999.9M:
0.1:
(1%+30)
Around 2.8V
Open at less than 30:
220V rms/10 seconds
AD Instruments
AC Voltage
30V~750V (50/60Hz)
1000V
2M:a99.9M:
100~999M:
1.00~10.00G:
DC1000V + 20%, -0%
1mA~1.2mA @1M:
26