USER MANUAL 400Amp True RMS AC/DC Clamp Meter Model MA435T Additional User Manual Translations available at www.extech.com GlobalTestSupply www. .com nd Quality Products Online at: sales@GlobalTestSupply.co...
Page 2
Introduction Congratulations on your purchase of this Extech MA435T True RMS Clamp Meter. This meter measures AC/DC Current, AC/DC Voltage, Resistance, Capacitance, Frequency, Diode Test, Duty Cycle and Continuity. Special features include Thermocouple Temperature and Non-Contact Voltage detection. The double molded case is designed for heavy duty use. This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.
SAFETY NOTES Do not exceed the maximum allowable input range of any function. Do not apply voltage to meter when resistance function is selected. Set the function switch OFF when the meter is not in use. Remove the battery if meter is to be stored for longer than 60 days.
Page 4
Description Meter Description NCV sensor Current clamp Non-contact AC voltage indicator light Clamp trigger Data Hold/ Back Light button LCD display MODE select button COM input jack Rotary Function switch 10. Relative button 11. Range button 12. Hz% Hold button 13.
Operation NOTES: Read and understand all Warning and Caution statements in this operation manual prior to using this meter. Set the function select switch to the OFF position when the meter is not in use. Non-Contact Voltage Detector WARNING: Risk of Electrocution. Before use, always test the Voltage Detector on a known live circuit to verify proper operation.
Page 6
AC/DC Voltage, Frequency, Duty Cycle Measurements CAUTION: Do not measure voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that can damage the meter. Rotate the function switch to the V position. Press the MODE button to select AC or DC Voltage.
Page 7
Capacitance Measurements WARNING: To avoid electric shock, discharge the capacitor before measuring. Rotate the function switch to the CAP capacitance position. Insert the black test lead banana plug into the negative COM jack. Insert the red test lead banana plug into the positive CAP jack. Touch the black test probe tip to one side of the device.
Page 8
Data Hold To freeze the LCD reading, press the HOLD button. The HOLD icon appears on the LCD. Press the HOLD button again to return to normal operation. Relative Press the REL button (REL will appear in the display) to zero the reading and create a reference point.
Maintenance WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the test leads from the input terminals, and turn OFF the meter before opening the case. Do not operate the meter with an open case. Cleaning and Storage Periodically wipe the case with a damp cloth and mild detergent;...
Specifications Accuracy Function Range Resolution (% of reading + digits) 40.00 A 0.01A AC Current* ±(2.8% + 8 digits) 50/60Hz TRMS 400.0 A 0.1A 40.00 A 0.01A ±(2.5% + 5 digits) DC Current 400.0 A 0.1A ±(2.8% + 5 digits) 400.0mV 0.1mV ±(1.5% + 30 digits)
Page 11
General Specifications 30mm (1.18") approx. Clamp jaw opening 4,000 count LCD Display Threshold <150; Test current < 0.5mA Continuity check Test current of 0.3mA typical; Diode test Open circuit voltage 1.5VDC typical Battery symbol is displayed Low Battery indication ‘OL’...