CALIBRATION AND MAINTENANCE

There are no internal user adjustments. Calibration is performed
without opening the instrument.
Before commencing calibration:
Clean all optical connectors very carefully.
Ensure that all devices have been at a stable room
temperature for over an hour, and that the light source is fully
warmed up at the wavelength to be calibrated.
Ensure that all installed batteries are in good working
condition.
KI 9000 UM-14
To enable calibration mode, remove belt clip and anti-tamper label at
the back of the instrument and insert a 2.54 mm (0.1") pitch
programming shunt across pins. Manipulation of the shunt is easier
with needle nose pliers.
Figure1. Calibration opening, positioned at the back of the instrument
underneath belt clip and covered by anti-temper label.
Perform calibration as per instructions below. When calibration is
complete, remove the calibration shunt, and place an anti-tamper label
over the hole. Do not forget to update your calibration records and to
schedule the next calibration service.