1-9-13 Zero Execute/Cancel; 1-9-14 Software Reset - Omron K3HB-PPB Technical Communications User's Manual

Digital indicators
Table of Contents

Advertisement

Commands and Responses (CompoWay/F Communications)
Response

1-9-13 Zero Execute/Cancel

Command
Response

1-9-14 Software Reset

Command
Response
MRC
SRC
Response code
"30"
"05"
"0000"
Response codes: The above indicates a normal end. For the response codes,
see 1-7 Operation Commands (CompoWay/F Communications) (page 1-11).
Command
Related
MRC
SRC
code
information
"30"
"05"
"03"
Related information
"00"
Zero cancel
"01"
Zero execute
This command is used to execute/cancel the forced zero and tare zero func-
tions. The results of the command will depend on the status of the tare zero
setting, as shown below.
Tare
Commands executed in order left to right.
zero
Zero execute
OFF
Forced zero ON
ON
Forced zero ON
This command can be used only in setting area 0. An operation error will
occur in the following situations.
• When there is no-measurement status, there is an input error, or a mea-
surement is outside the display range (but zero cancel is possible.)
• When a RESET or HOLD input is ON
• When the command is used in setting area 1
MRC
SRC
Response code
"30"
"05"
"0000"
Response codes: The above indicates a normal end. For the response codes,
see 1-7 Operation Commands (CompoWay/F Communications) (page 1-11).
Command
Related
MRC
SRC
code
information
"30"
"05"
"06"
"00"
A software reset causes the same operation as turning the power off and on.
This command can be used in both setting area 0 and setting area 1.
To use this command, "Write via communications" must be enabled using the
"Write via communications" operation command.
(No response)
A response is not returned to this operation command.
Description
Zero execute
Zero cancel
Forced zero ON
Forced zero OFF Forced zero OFF
Tare zero ON
Tare zero OFF
Section 1-9
Zero cancel
Forced zero OFF
1-19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents