How to Measure Blood Pressure (in MANU. Mode)
5. Place the stethoscope on the patient's
arm.
6. Push the START Button to start the
measurement.
• Do not push the START Button without wrapping the cuff on
the arm.
If a pulse rate is detected during deflation, the pressure
display will be synchronized with the pulse rate and renewed.
• If you want to inflate again after the start of deflation, you
can inflate the monitor again by pushing the START Button
only while you are pushing it.
• If you want to accelerate deflation after the start of deflation,
you can accelerate it by pushing the DEFLATION (deflation
control) / Measurement Results Display Switch Button. Each
time the Button is pushed, the deflation speed can be
accelerated by approximately 5 to 10 mmHg.
7. Push the STOP Button to finish the
measurement.
• While the battery pack is in use, the monitor will turn off
automatically in approximately five minutes and the display
(measurement results) will disappear. (Automatic Power Off)
8. Push the ON/OFF (power) Button to
turn off the power.
If the monitor detects a noise during the
measurement or determines that the
pressure value is not correct, an error
display appears (Er1 to 9). In this case, refer
to Page 24 for the remedial actions and start
the measurement again.
ON/OFF
START
HIDE
P-SET
MODE
140
180
SINGLE
AVG.
MANU.
220
100
CHECK
STOP
260
AUTO
280
mmHg
ON/OFF
AC
START
HIDE
P - SET
MODE
SINGLE
140
180
MANU.
AVG.
220
CHECK
100
STOP
260
AUTO
280
mmHg
ON/OFF
START
HIDE
P - SET
MODE
SINGLE
140
180
AVG.
MANU.
220
CHECK
100
STOP
260
280
AUTO
ON/OFF
START
HIDE
P - SET
MODE
SINGLE
140
180
AVG.
MANU.
220
CHECK
100
STOP
260
280
AUTO
24

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hem-907

Table of Contents