Command: TMID
Purpose:
Response:
Command String:
Description:
Response String:
Description:
Example:
Manual 19854 Rev. D
ECN 50523
Writes a number (0 to 511) to the measurement position ID field of a TEDS
sensor.
The module responds with a "0" to tell the calling process the command has
been received.
XYC0ZTMID
X—Rack number of the target module (Range: ASCII printable characters, 0
through 3)
Y— Slot number of the target module. (Range: ASCII printable characters, 0
through 7)
C0Z— Module type for the target. Specifying "C01" in this argument addresses
a 443B101, specifying "C02" addresses a 443B102.
TMID—tells the module to write to the measurement position ID field of a TEDS
sensor.
"0"
This return indicates that the command was received by the card.
Command String:
"06C02TMID132"
Response String:
"0"
The 443B102 in rack 0 and slot 6 is commanded to write the number 132
to the measurement position ID field of the TEDS sensor.
443B101/443B102
37