PCB Piezotronics 443B02 Installation And Operating Manual page 33

Modular ac line power supply
Table of Contents

Advertisement

Command: XDCR
Purpose:
Response:
Command String:
Description:
Response String:
Description:
Example:
Manual 19854 Rev. D
ECN 50523
Sets the transducer sensitivity of the amplifier.
The module responds with a "0" to tell the calling process the command has
been received.
XYCOZXDCRXX.XX
X— Rack number of the target module (Range: ASCII printable characters, 0
through 3)
Y— Slot number of the target module. (Range: ASCII printable characters, 0
through 7)
C0Z— Module type for the target. Specifying "C01" in this argument addresses
a 443B101, specifying "C02" addresses a 443B102.
XDCR— Commands the module to set the output sensitivity
XX.XX— 5 character transducer sensitivity. (Range: ASCII printable characters, 0
through 9)
"0"
This return indicates that the command was received by the card.
Command String:
"04C01XDCR2.034"
Response String:
"0"
The output sensitivity of the 443B101 in rack 0, and slot 4 will be set to 2.034.
443B101/443B102
13

Advertisement

Table of Contents
loading

This manual is also suitable for:

441a33443b102441a101441a35441a42441a43 ... Show all

Table of Contents