Command: LOWF
Purpose:
Response:
Command String:
Description:
Response String:
Description:
Example:
Manual 19854 Rev. D
ECN 50523
Sets the low frequency response of the amplifier.
The module responds with a "0" to tell the calling process the command has
been received.
XYC0ZLOWFZ
X—Rack number of the target module (Range: ASCII printable characters, 0
through 3)
Y— Slot number of the target module. (Range: ASCII printable characters, 0
through 7)
C0Z— Module type for the target. Specifying "C01" in this argument addresses
a 443B101, specifying "C02" addresses a 443B102.
LOWF— Commands module to change the low frequency response.
Z— specifies the low frequency response as follows:
"1" sets the low frequency response to 0.2 Hz.
"2" sets the low frequency response to 2.0 Hz
"3" engages the medium time constant.
"4" engages the long time constant.
"0"
This return indicates that the command was received by the card.
Command String:
"06C02LOWF3"
Response String:
"0"
The 443B102 in rack 0 and slot 6 is set to medium time constant mode.
443B101/443B102
17